This paper describes a detailed protocol for obtaining chronic, multi-site unit recordings in cerebral cortex of awake animals for periods of three months or more. The protocol includes details for malting relatively simple and inexpensive implantable multichannel electrodes that consist of arrays of separate microwires. The results reported in this paper suggest that a viable implant will have discriminable unit activity on about 80% of the electrodes, resulting in, on average, the simultaneous unit recording of upwards of 60 units during a daily recording session. The active electrodes during one recording session tend to remain active in subsequent recording sessions for several weeks. Using the methods described here, implants have been constructed which incorporate several different electrode materials, coatings, sizes, and electrode separation within a single array. These microwire electrode arrays provide the basic technology fur obtaining unit recordings for several months. This provides a model system for studying biocompatibility of neural implants, which is a critical component for the development of neural implants that have an indefinite working span. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:303 / 313
页数:11
相关论文
共 24 条
[1]   HISTOPATHOLOGIC EVALUATION OF PROLONGED INTRACORTICAL ELECTRICAL-STIMULATION [J].
AGNEW, WF ;
YUEN, TGH ;
MCCREERY, DB ;
BULLARA, LA .
EXPERIMENTAL NEUROLOGY, 1986, 92 (01) :162-185
[2]  
Chapin J. K., 1998, Society for Neuroscience Abstracts, V24, P658
[3]   Functional connectivity in auditory cortex using chronic, multichannel unit recordings [J].
Clement, RS ;
Witte, RS ;
Rousche, PJ ;
Kipke, DR .
NEUROCOMPUTING, 1999, 26-7 :347-354
[4]   PERFORMANCE OF PLANAR MULTISITE MICROPROBES IN RECORDING EXTRACELLULAR SINGLE-UNIT INTRACORTICAL ACTIVITY [J].
DRAKE, KL ;
WISE, KD ;
FARRAYE, J ;
ANDERSON, DJ ;
BEMENT, SL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (09) :719-732
[5]   Tetrodes markedly improve the reliability and yield of multiple single-unit isolation from multi-unit recordings in cat striate cortex [J].
Gray, CM ;
Maldonado, PE ;
Wilson, M ;
McNaughton, B .
JOURNAL OF NEUROSCIENCE METHODS, 1995, 63 (1-2) :43-54
[6]   A 3-DIMENSIONAL MICROELECTRODE ARRAY FOR CHRONIC NEURAL RECORDING [J].
HOOGERWERF, AC ;
WISE, KD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (12) :1136-1146
[7]   A GLASS SILICON COMPOSITE INTRACORTICAL ELECTRODE ARRAY [J].
JONES, KE ;
CAMPBELL, PK ;
NORMANN, RA .
ANNALS OF BIOMEDICAL ENGINEERING, 1992, 20 (04) :423-437
[8]   Plasma-etched neural probes [J].
Kewley, DT ;
Hills, MD ;
Borkholder, DA ;
Opris, IE ;
Maluf, NI ;
Storment, CW ;
Bower, JM ;
Kovacs, GTA .
SENSORS AND ACTUATORS A-PHYSICAL, 1997, 58 (01) :27-35
[9]  
KIPKE DR, 1996, ANN BIOMED ENG, V24, P68
[10]  
Maynard E. M., 1996, Society for Neuroscience Abstracts, V22, P1829