Mediator-free amperometric determination of glucose based on direct electron transfer between glucose oxidase and an oxidized boron-doped diamond electrode

被引:34
作者
Jing Wu [1 ]
Yang Qu [1 ]
机构
[1] Huazhong Agr Univ, Coll Sci, Dept Chem, Wuhan 430070, Peoples R China
关键词
biosensor; glucose; glucose oxidase (GOD); oxidized boron-doped diamond electrode;
D O I
10.1007/s00216-006-0534-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A mediator-free glucose biosensor, termed a "third-generation biosensor," was fabricated by immobilizing glucose oxidase (GOD) directly onto an oxidized boron-doped diamond (BDD) electrode. The surface of the oxidized BDD electrode possesses carboxyl groups (as shown by Raman spectra) which covalently cross-link with GOD through glutaraldehyde. Glucose was determined in the absence of a mediator used to transfer electrons between the electrode and enzyme. O-2 has no effect on the electron transfer. The effects of experimental variables (applied potential, pH and cross-link time) were investigated in order to optimize the analytical performance of the amperometric detection method. The resulting biosensor exhibited fast amperometric response (less than 5 s) to glucose. The biosensor provided a linear response to glucose over the range 6.67 x 10(-5) to 2 x 10(-3) mol/L, with a detection limit of 2.31 x 10(-5) mol/L. The lifetime, reproducibility and measurement repeatability were evaluated and satisfactory results were obtained.
引用
收藏
页码:1330 / 1335
页数:6
相关论文
共 44 条
[1]   Electrochemical sensors [J].
Bakker, E .
ANALYTICAL CHEMISTRY, 2004, 76 (12) :3285-3298
[2]   Electrical communication between glucose oxidase and electrodes mediated by phenothiazine-labeled poly(ethylene oxide) bonded to lysine residues on the enzyme surface [J].
Ban, K ;
Ueki, T ;
Tamada, Y ;
Saito, T ;
Imabayashi, S ;
Watanabe, M .
ANALYTICAL CHEMISTRY, 2003, 75 (04) :910-917
[3]   ELECTROCHEMICAL IMMOBILIZATION OF ENZYMES .3. IMMOBILIZATION OF GLUCOSE-OXIDASE IN THIN-FILMS OF ELECTROCHEMICALLY POLYMERIZED PHENOLS [J].
BARTLETT, PN ;
TEBBUTT, P ;
TYRRELL, CH .
ANALYTICAL CHEMISTRY, 1992, 64 (02) :138-142
[4]   Layer-by-layer self-assembly of glucose oxidase and Os(Bpy)2CIPyCH2NH-poly(allylamine) bioelectrode [J].
Calvo, EJ ;
Etchenique, R ;
Pietrasanta, L ;
Wolosiuk, A ;
Danilowicz, C .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1161-1168
[5]   The electrooxidation of sulfur-containing compounds at boron-doped diamond electrode [J].
Chailapakul, O ;
Aksharanandana, P ;
Frelink, T ;
Einaga, Y ;
Fujishima, A .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 80 (03) :193-201
[6]   Electrochemically mediated electrodeposition/electropolymerization to yield a glucose microbiosensor with improved characteristics [J].
Chen, XH ;
Matsumoto, N ;
Hu, YB ;
Wilson, GS .
ANALYTICAL CHEMISTRY, 2002, 74 (02) :368-372
[7]   A rapid and easy procedure of biosensor fabrication by microencapsulation of enzyme in hydrophilic synthetic latex films. Application to the amperometric determination of glucose [J].
Cosnier, S ;
Szunerits, S ;
Marks, RS ;
Novoa, A ;
Puech, L ;
Perez, E ;
Rico-Lattes, I .
ELECTROCHEMISTRY COMMUNICATIONS, 2000, 2 (12) :851-855
[8]  
CSOREGI E, 1995, ANAL CHEM, V67, P1240
[9]   Electrochemical performance of diamond thin-film electrodes from different commercial sources [J].
Fischer, AE ;
Show, Y ;
Swain, GM .
ANALYTICAL CHEMISTRY, 2004, 76 (09) :2553-2560
[10]  
FORTIN E, 2006, SCANNIGN ELECTROCHEM