A comparison between different methods for the determination of reduced and oxidized glutathione in mammalian tissues

被引:115
作者
Floreani, M
Petrone, M
Debetto, P
Palatini, P
机构
[1] Department of Pharmacology, University of Padova, 35131 Padova
关键词
glutathione assay; fluorometric; enzymatic; colorimetric; HPLC methods;
D O I
10.3109/10715769709084481
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
this study, three rapid assay techniques for the determination of glutathione, one enzymatic, one fluorometric and one newly patented colorimetric method, were compared by measuring reduced (GSH) and oxidized (GSSG) glutathione in guineapig heart and liver. The HPLC technique was used as a standard, since it is considered the most reliable assay method. In heart, all methods measured the same levels of GSH (about 1 mu mole/g wet tissue), whereas in liver the fluorometric assay gave GSH levels about half as high as those measured by the other methods (about 3 vs. 7 mu moles/g wet tissue). Conversely, the fluorometric assay grossly overestimated GSSG concentration (by 5 to 8 times) in both heart and liver. These results confirm previous doubts about the use of the fluorometric technique for GSSG determination in mammalian tissues and also raise some questions about its use for the measurement of GSH in liver. In this tissue, the GSH concentration determined by the fluorometric method was shown to be inversely correlated with the size of the sample, suggesting the presence of some quenching material.