Improvement of surface acoustic wave gas and biosensor response characteristics using a capacitive coupling technique

被引:30
作者
Bender, F [1 ]
Länge, K [1 ]
Voigt, A [1 ]
Rapp, M [1 ]
机构
[1] Forschungszentrum Karlsruhe, Inst Instrumental Anal, D-76021 Karlsruhe, Germany
关键词
D O I
10.1021/ac035019+
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
As for many other electronic devices and circuits, electrical contact to surface acoustic wave sensors is usually made using bonding wires. Ibis technique is known to result in reliable contact under most conditions, but it does so with several disadvantages. First, electrical contact is not reversible, impeding replacement of sensor devices. Second, bonding wires are quite delicate and should not be exposed to high gas or liquid flows. Third, the presence of bonding wires may limit the potential to miniaturize a sensor housing or flow cell. Therefore, a capacitive coupling technique was developed to replace bonding wires. This permitted redesign of flow cells and sensor arrays, resulting in flow cell volumes of 80 muL to 60 nL. As a consequence, response times were reduced to 1-2 s in gas sensing and a few minutes in liquid sensing, respectively. At the same time, sensor devices can be easily replaced, and the system is less susceptible to malfunction.
引用
收藏
页码:3837 / 3840
页数:4
相关论文
共 14 条
[11]   A dual-adsorbent preconcentrator for a portable indoor-VOC microsensor system [J].
Lu, CJ ;
Zellers, ET .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3449-3457
[12]   New miniaturized SAW-sensor array for organic gas detection driven by multiplexed oscillators [J].
Rapp, M ;
Reibel, J ;
Voigt, A ;
Balzer, M ;
Bülow, O .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 65 (1-3) :169-172
[13]  
RAPP M, 2002, 202 M EL SOC SALT LA
[14]   DIRECT PIEZOELECTRIC COUPLING TO SURFACE ELASTIC WAVES [J].
WHITE, RM ;
VOLTMER, FW .
APPLIED PHYSICS LETTERS, 1965, 7 (12) :314-&