The design of an animal PET: Flexible geometry for achieving optimal spatial resolution or high sensitivity

被引:51
作者
Weber, S [1 ]
Terstegge, A [1 ]
Herzog, H [1 ]
Reinartz, R [1 ]
Reinhart, P [1 ]
Rongen, F [1 ]
MullerGartner, HW [1 ]
Halling, H [1 ]
机构
[1] FORSCHUNGSZENTRUM JULICH, FORSCHUNGSZENTRUM, ZENT LAB ELEKT, D-52425 JULICH, GERMANY
关键词
animal PET; PET instrumentation; simulation; YAP;
D O I
10.1109/42.640759
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We present the design of a positron emission tomograph (PET) with flexible geometry dedicated to in who studies of small animals (TierPET). The scanner uses two pairs of detectors, Each detector consists of 400 small individual yttrium aluminum perovskite (PAP) scintillator crystals of dimensions 2x2x15 mm(3), optically isolated and glued together, which are coupled to position-sensitive photomultiplier tubes (PSPMT's). The detector modules can be moved in a radial direction so that the detector-to-detector spacing can be varied, Special hardware has been built for coincidence detection, position detection, and real-time data acquisition, which is performed by a PC. The single-event data are transferred to workstations where the radioactivity distribution is reconstructed. The dimensions of the crystals and the detector layout are the result of extensive simulations which are described in this report, taking into account sensitivity, spatial resolution and additional parameters like parallax error or scatter effects, For the three-dimensional (3-D) reconstruction a genuine 3-D expectation-maximization (EM)-algorithm which can include the characteristics of the detector system has been implemented, The reconstruction software is flexible and matches the different detector configurations, The main advantage of the proposed animal PET scanner is its high flexibility, allowing the realization of various detector-system configurations. By changing the detector-to-detector spacing, the system is capable of either providing good spatial resolution or high sensitivity for dynamic studies of pharmacokinetics.
引用
收藏
页码:684 / 689
页数:6
相关论文
共 15 条
[1]   SCINTILLATION PROPERTIES OF YAP-CE [J].
BACCARO, S ;
BLAZEK, K ;
DENOTARISTEFANI, F ;
MALY, P ;
MARES, JA ;
PANI, R ;
PELLEGRINI, R ;
SOLURI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 361 (1-2) :209-215
[2]   THE DESIGN AND PHYSICAL CHARACTERISTICS OF A SMALL ANIMAL POSITRON EMISSION TOMOGRAPH [J].
BLOOMFIELD, PM ;
RAJESWARAN, S ;
SPINKS, TJ ;
HUME, SP ;
MYERS, R ;
ASHWORTH, S ;
CLIFFORD, KM ;
JONES, WF ;
BYARS, LG ;
YOUNG, J ;
ANDREACO, M ;
WILLIAMS, CW ;
LAMMERTSMA, AA ;
JONES, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) :1105-1126
[3]  
*CERN, 1994, GEANT DET DESCR SIM
[4]  
CUTLER PD, 1992, J NUCL MED, V33, P595
[5]   EVALUATION OF CERIUM-DOPED LUTETIUM OXYORTHOSILICATE (LSO) SCINTILLATION CRYSTAL FOR PET [J].
DAGHIGHIAN, F ;
SHENDEROV, P ;
PENTLOW, KS ;
GRAHAM, MC ;
ESHAGHIAN, B ;
MELCHER, CL ;
SCHWEITZER, JS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :1045-1047
[6]   LIGHT COLLECTION IN SCINTILLATION DETECTOR COMPOSITES FOR NEUTRON DETECTION [J].
KNOLL, GF ;
KNOLL, TF ;
HENDERSON, TM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :872-875
[7]   TWO-DIMENSIONAL NEUTRON DETECTOR BASED ON A POSITION-SENSITIVE PHOTOMULTIPLIER [J].
KURZ, R ;
REINARTZ, R ;
WIDDAU, S ;
SCHELTEN, J ;
SCHOLZ, A ;
SCHAFER, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 273 (01) :273-282
[8]   DESIGN AND ENGINEERING ASPECTS OF A HIGH-RESOLUTION POSITRON TOMOGRAPH FOR SMALL ANIMAL IMAGING [J].
LECOMTE, R ;
CADORETTE, J ;
RICHARD, P ;
RODRIGUE, S ;
ROULEAU, D .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1446-1452
[9]  
LEMPICKI A, 1994, C REC IEEE NUCL SCI, P307
[10]   PERFORMANCE OF QPET, A HIGH-RESOLUTION 3D PET IMAGING-SYSTEM FOR SMALL VOLUMES [J].
MCKEE, BTA ;
DICKSON, AW ;
HOWSE, DC .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) :176-185