Assessments of different kinds of stems by experiments and FEM analysis: Appropriate stress distribution on a hip prosthesis

被引:34
作者
Sakai, Rina
Roman, Moritoshi
Mabuchi, Kiyoshi
机构
[1] Kitasato Univ, Sch Allied Hlth Sci, Dept Biomed Engn, Sagamihara, Kanagawa 2288555, Japan
[2] Kitasato Univ, Sch Med, Dept Orthoped Surg, Sagamihara, Kanagawa 2288555, Japan
关键词
prosthetic stem; fixation method; finite element method; loading test; stress distribution;
D O I
10.1016/j.clinbiomech.2006.03.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Background. It is now recognized that initial stability is essential for avoidance of thigh pain in hip replacement. The initial stability corresponds to an optimal stress distribution of cementless orthopedic implants. Although the relationship between the contour and stress at the fixation site has been analyzed, guidelines on stem design have not been established. Methods. Finite element models of three currently-used stems were constructed for a computer simulation. Contact stress at the fixation site of a joint prosthesis was analyzed by an explicit three-dimensional finite element method. The stress immediately after applying load using a film or sensor which can measure contact stress was observed. The situation of the initial fixation about the specific part which becomes important clinically based on the results was clarified. Findings. We introduced fluctuation area as a measure to evaluate the primary fixation of femoral stems. It was found that the stress distribution on the PerFix SV stem fluctuated with a slight disturbance. On the Intra-Medullary Cruciate stem, the high stress areas were distributed on the proximal area and under the pin. The high stress area on the VerSys stem were spread on the medial side. Interpretation. This study highlights the mechanical instability of the fixation site of joint prostheses, and thus suggests that the general idea that unconditionally recommends a larger area for the fixation site of joint prostheses should be revised. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:826 / 833
页数:8
相关论文
共 26 条
[1]
AMAO K, 2003, J JAPANESE SOC CLIN, V24, P293
[2]
Three-dimensional primary stability of cementless femoral stems [J].
Buhler, DW ;
Berlemann, U ;
Lippuner, K ;
Jaeger, P ;
Nolte, LP .
CLINICAL BIOMECHANICS, 1997, 12 (02) :75-86
[3]
A three-dimensional finite element model from computed tomography data: a semi-automated method [J].
Cattaneo, PM ;
Dalstra, M ;
Frich, LH .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H2) :203-213
[4]
Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs [J].
Cristofolini, L ;
Teutonico, AS ;
Monti, L ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1603-1615
[5]
In vivo determination of hip joint separation and the forces generated due to impact loading conditions [J].
Dennis, DA ;
Komistek, RD ;
Northcut, EJ ;
Ochoa, JA ;
Ritchie, A .
JOURNAL OF BIOMECHANICS, 2001, 34 (05) :623-629
[6]
Primary stability in cementless femoral stems:: custom-made versus conventional femoral prosthesis [J].
Götze, C ;
Steens, W ;
Vieth, V ;
Poremba, C ;
Claes, L ;
Steinbeck, J .
CLINICAL BIOMECHANICS, 2002, 17 (04) :267-273
[7]
CORRELATIONS BETWEEN ORTHOGONAL MECHANICAL-PROPERTIES AND DENSITY OF TRABECULAR BONE - USE OF DIFFERENT DENSITOMETRIC MEASURES [J].
KEYAK, JH ;
LEE, IY ;
SKINNER, HB .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (11) :1329-1336
[8]
Influence of preoperative mechanical bone quality and bone mineral density on aseptic loosening of total hip arthroplasty after seven years [J].
Krischak, GD ;
Wachter, NJ ;
Zabel, T ;
Suger, G ;
Beck, A ;
Kinzl, L ;
Claes, LE ;
Augat, P .
CLINICAL BIOMECHANICS, 2003, 18 (10) :916-923
[9]
Thigh pain in primary total hip arthroplasty - The effects of elastic moduli [J].
Lavernia, C ;
D'Apuzzo, M ;
Hernandez, V ;
Lee, D .
JOURNAL OF ARTHROPLASTY, 2004, 19 (07) :10-16
[10]
Measurement of the migration of a cemented hip prosthesis in an in vitro test [J].
Maher, SA ;
Prendergast, PJ ;
Lyons, CG .
CLINICAL BIOMECHANICS, 2001, 16 (04) :307-314