Metal artifact reduction in musculoskeletal magnetic resonance imaging

被引:58
作者
Harris, Craig A.
White, Lawrence M.
机构
[1] Univ Toronto, Mt Sinai Hosp, Div Musculoskeletal Imaging, Toronto, ON M5G 1X5, Canada
[2] Univ Hlth Network, Toronto, ON M5G 1X5, Canada
关键词
D O I
10.1016/j.ocl.2006.04.001
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Many strategies can be employed to reduce the size and scale of metal susceptibility artifacts in the vicinity of orthopedic hardware; factors include selection of metal hardware material, patient positioning, and MRI sequence adjustments and techniques. The adjustments to sequence parameters include high spatial resolution fast spin echo sequences with minimal interecho spacing and increased receiver bandwidth. Frequency and phase encoding gradients can be orientated so misregistration artifacts arising from the metal are directed away from areas of anticipated clinical diagnostic interest. Complications arising in the vicinity of metallic hardware, including loosening, can be assessed after implication of these metal artifact reduction techniques.
引用
收藏
页码:349 / +
页数:12
相关论文
共 47 条
[1]
SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING [J].
BAKKER, CJG ;
BHAGWANDIEN, R ;
MOERLAND, MA ;
RAMOS, LMP .
MAGNETIC RESONANCE IMAGING, 1994, 12 (05) :767-774
[2]
SUSCEPTIBILITY ARTIFACTS IN 2DFT SPIN-ECHO AND GRADIENT-ECHO IMAGING - THE CYLINDER MODEL REVISITED [J].
BAKKER, CJG ;
BHAGWANDIEN, R ;
MOERLAND, MA ;
FUDERER, M .
MAGNETIC RESONANCE IMAGING, 1993, 11 (04) :539-548
[3]
MR-IMAGING IN EVALUATION OF SUSPECTED HIP FRACTURE - FREQUENCY OF UNSUSPECTED BONE AND SOFT-TISSUE INJURY [J].
BOGOST, GA ;
LIZERBRAM, EK ;
CRUES, JV .
RADIOLOGY, 1995, 197 (01) :263-267
[4]
Burtscher IM, 1998, ACTA RADIOL, V39, P70
[5]
MRI of spinal hardware: comparison of conventional T1-weighted sequence with a new metal artifact reduction sequence [J].
Chang, SD ;
Lee, MJ ;
Munk, PL ;
Janzen, DL ;
MacKay, A ;
Xiang, QS .
SKELETAL RADIOLOGY, 2001, 30 (04) :213-218
[6]
TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[7]
Czerny C, 1999, Top Magn Reson Imaging, V10, P214, DOI 10.1097/00002142-199908000-00003
[8]
OCCULT FRACTURES OF THE PROXIMAL FEMUR - MR IMAGING [J].
DEUTSCH, AL ;
MINK, JH ;
WAXMAN, AD .
RADIOLOGY, 1989, 170 (01) :113-116
[9]
A comparison of conventional spin-echo and turbo spin-echo imaging of soft tissues adjacent to orthopedic hardware [J].
Eustace, S ;
Jara, H ;
Goldberg, R ;
Fenlon, H ;
Mason, M ;
Melhem, ER ;
Yucel, EK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 170 (02) :455-458
[10]
FELDMAN F, 1994, SKELETAL RADIOL, V23, P439