Relative profile and dose verification of intensity-modulated radiation therapy

被引:60
作者
Chang, JW [1 ]
Mageras, GS [1 ]
Chui, CS [1 ]
Ling, CC [1 ]
Lutz, W [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2000年 / 47卷 / 01期
关键词
EPID; IMRT; treatment QA; dose verification; treatment field verification;
D O I
10.1016/S0360-3016(99)00555-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop a quality assurance (QA) procedure to assess the intensity profile and dosimetry for intensity-modulated (IM) treatment fields using electronic portal imaging devices (EPIDs), Methods and Materials: A series of rapidly acquired (similar to 1/sec) portal images are summed and converted to dose, For relative intensity QA, the intended profile is subtracted point-by-point from the measured profile forming a series of error values. The standard deviation, a, of the errors, a measure of the goodness of the match, is minimized by applying a normalization and uniform scatter subtraction from the measured profile. For dose verification (dose to isocenter), an empirically determined phantom-correction factor is added to incorporate the effect of patient presence on EPID readings. Seventy prostate treatment fields were used in a phantom study to verify these approaches, Sensitivity was studied by creating artificial mismatches. Results: The average a for relative profile verification is 3.3% (percentage of average intended intensity) whereas artificial mismatches resulted in a values from 5% to 27%, The average isocentric dose calculated from EPID readings is 1.001 relative to the planned dose with a standard deviation of 0.018. Conclusions: An EPID can be used for profile verification and absolute isocentric dose measurement for IM fields. (C) 2000 Elsevier Science Inc.
引用
收藏
页码:231 / 240
页数:10
相关论文
共 24 条
[1]   Specification of dose delivery in radiation therapy - Recommendations by the Nordic Association of Clinical Physics (NACP) [J].
Aaltonen, P ;
Brahme, A ;
Lax, I ;
Levernes, S ;
Naslund, I ;
Reitan, JB ;
Turesson, I .
ACTA ONCOLOGICA, 1997, 36 :1-32
[2]   The dose response relationship of a liquid-filled electronic portal imaging device [J].
Boellaard, R ;
vanHerk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 1996, 23 (09) :1601-1611
[3]   A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS) [J].
BOYER, AL ;
ANTONUK, L ;
FENSTER, A ;
VANHERK, M ;
MEERTENS, H ;
MUNRO, P ;
REINSTEIN, LE ;
WONG, J .
MEDICAL PHYSICS, 1992, 19 (01) :1-16
[4]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[5]  
CHANG J, 1999, MED PHYS, V26, P1166
[6]  
CHANG J, 1999, MED PHYS, V26, P1168
[7]   DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS [J].
CHUI, CS ;
LOSASSO, T ;
SPIROU, S .
MEDICAL PHYSICS, 1994, 21 (08) :1237-1244
[8]   Transmission dosimetry with a liquid-filled electronic portal imaging device [J].
Essers, M ;
Boellaard, R ;
vanHerk, M ;
Lanson, H ;
Mijnheer, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (04) :931-941
[9]   Dosimetric characteristics of a liquid-filled electronic portal imaging device [J].
Essers, M ;
Hoogervorst, BR ;
vanHerk, M ;
Lanson, H ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1265-1272
[10]   X-RAY SOURCES OF MEDICAL LINEAR ACCELERATORS - FOCAL AND EXTRA-FOCAL RADIATION [J].
JAFFRAY, DA ;
BATTISTA, JJ ;
FENSTER, A ;
MUNRO, P .
MEDICAL PHYSICS, 1993, 20 (05) :1417-1427