Electrospray LC-MS/MS quantitation, stability, and preliminary pharmacokinetics of bradykinin antagonist polypeptide B201 (NSC 710295) in the mouse

被引:20
作者
Feng, WY
Chan, KK [1 ]
Covey, JM
机构
[1] Ohio State Univ, Coll Pharm, Columbus, OH 43210 USA
[2] Ohio State Univ, Coll Med, Columbus, OH 43210 USA
[3] Natl Canc Inst, Rockville, MD 20852 USA
关键词
LC-MS/MS assay; ion-trap mass spectrometry; B201; polypeptide; bradykinin antagonist;
D O I
10.1016/S0731-7085(01)00697-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
B201 (NSC 710295), [SUIM-(Darg-Arg-Pro-Hyp-Gly-Igl-Ser-Digl-Oic-Arg)(2)], a third generation of bradykinin (BK) antagonist, has been found to possess high potency. We report the development of a highly sensitive electrospray LC-MS/MS assay method for the analysis of B201 in plasma for the first time, using an ion-trap mass spectrometer. Human or mouse plasma (0.2 ml) was spiked with B201 and the internal standard, substance P. The compounds were extracted with a preconditioned C-18 reversed-phase column and analyzed by LC-MSJMS. The analytes were separated on a 50 x 2 mm (i.d.) BetaBasic C8 column, using a gradient elution. The positive ion selected reaction monitor mode was used monitoring the transitions of ions at m/z 938.9(3+) --> 816.0(2+) for B201 and 674.3(2+) --> 665.7(2+) for substance P. Assay validation was performed, and the limit of quantitation (LOQ) for B201 was found to be 1 ng/ml for human plasma and 2.5 ng/ml for mouse plasma. The recovery was 78% for B201 and 88% for substance P. The assay was linear from 2.5 to 1500 ng/ml for mouse plasma monitored. Using a 0.2 ml plasma, the within-day CVs were 9.3% at 2.5 ng/ml, 6.5% at 100 ng/ml, and 3.8% at 1000 ng/ml for human plasma (n = 6). For mouse plasma, the respective within-day CVs were 17.6, 9.6, and 6.2% (n = 6). The between-day CVs for human plasma were 8.2, 10.9, and 2.4%, respectively, (n = 3) and the respective values for mouse plasma were 11.9, 8.6 and 6.5% (n = 6). Pharmacokinetics of B201 in the mouse was studied following i.v. administration at 5 ma/kg and found to conform to a two-compartment model with an initial half-life of 14 min and a terminal half-life of 44 h. Plasma B201 peak level was detected at mug/ml range and the levels were detectable for a least 24 h. Preliminary oral bioavailability was found to be about 1%. This method demonstrates that an ion trap mass spectrometer can be a powerful tool to quantify large peptides at low nanogram per mililiter with a non-isotopically labeled internal standard, (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:601 / 612
页数:12
相关论文
共 27 条
[1]  
Bartolucci G, 2000, RAPID COMMUN MASS SP, V14, P967, DOI 10.1002/(SICI)1097-0231(20000615)14:11<967::AID-RCM973>3.0.CO
[2]  
2-C
[3]   Quantitative electrospray LC-MS and LC-MS/MS in biomedicine [J].
Carrascal, M ;
Schneider, K ;
Calaf, RE ;
van Leeuwen, S ;
Canosa, D ;
Gelpí, E ;
Abian, J .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1998, 17 (6-7) :1129-1138
[4]   A NEW CLASS OF BRADYKININ ANTAGONISTS - SYNTHESIS AND INVITRO ACTIVITY OF BISSUCCINIMIDOALKANE PEPTIDE DIMERS [J].
CHERONIS, JC ;
WHALLEY, ET ;
NGUYEN, KT ;
EUBANKS, SR ;
ALLEN, LG ;
DUGGAN, MJ ;
LOY, SD ;
BONHAM, KA ;
BLODGETT, JK .
JOURNAL OF MEDICINAL CHEMISTRY, 1992, 35 (09) :1563-1572
[5]   Detection and quantitation of cellularly derived amyloid β peptides by immunoprecipitation-HPLC-MS [J].
Clarke, NJ ;
Tomlinson, AJ ;
Ohyagi, Y ;
Younkin, S ;
Naylor, S .
FEBS LETTERS, 1998, 430 (03) :419-423
[6]   A COMPARISON BETWEEN ION TRAP AND TRIPLE QUADRUPOLE COLLISION-RESULTS [J].
EVANS, C ;
TRALDI, P ;
MELE, A ;
SOTTANI, C .
ORGANIC MASS SPECTROMETRY, 1991, 26 (04) :347-349
[7]   High sensitivity identification of proteins by electrospray ionization tandem mass spectrometry: Initial comparison between an ion trap mass spectrometer and a triple quadrupole mass spectrometer [J].
Figeys, D ;
Aebersold, R .
ELECTROPHORESIS, 1997, 18 (3-4) :360-368
[8]  
HOCK FJ, 1991, BRIT J PHARMACOL, V102, P769, DOI 10.1111/j.1476-5381.1991.tb12248.x
[9]   Rapid target analysis of microcontaminants in water by on-line single-short-column liquid chromatography combined with atmospheric pressure chemical ionization ion-trap mass spectrometry [J].
Hogenboom, AC ;
Niessen, WMA ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :201-210
[10]   TANDEM-IN-SPACE AND TANDEM-IN-TIME MASS-SPECTROMETRY - TRIPLE QUADRUPOLES AND QUADRUPOLE ION TRAPS [J].
JOHNSON, JV ;
YOST, RA ;
KELLEY, PE ;
BRADFORD, DC .
ANALYTICAL CHEMISTRY, 1990, 62 (20) :2162-2172