Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone-beam CT from kV on-board imagers

被引:22
作者
Ali, Imad [1 ]
Ahmad, Salahuddin [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, Oklahoma City, OK 73104 USA
关键词
sagging shifts; cone-beam CT (CBCT); on-board imager (OBI); isocenter; image quality; FLAT-PANEL IMAGER; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; RADIOSURGERY; ASSURANCE;
D O I
10.1120/jacmp.v10i3.2930
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
To investigate the effects of sagging shifts of three on-board kV imaging systems (OBI) on the isocenter positioning accuracy and image quality of cone-beam CT (CBCT). A cubical phantom having a metal marker in the center that can be aligned with the radiation isocenter was used to measure sagging shifts and their variation with gantry angle on three Varian linacs with kV on-board imaging systems. A marker-tracking algorithm was applied to detect the shadow of the metal marker and localize its center in the two-dimensional cone-beam radiographic projections. This tracking algorithm is based on finding the position of maximum cross-correlation between a region-of-interest from a template image (including the metal marker) and the projections containing the shadow of the metal marker. Sagging shifts were corrected by mapping the center of the metal marker to a reference position for all projections acquired over a full gantry rotation (0-360 degrees). The sag-corrected radiographic projections were then used to reconstruct CBCT using Feldkamp back-projection. A standard quality assurance phantom was used to evaluate the image quality of CBCT before and after sagging correction. Sagging affects both the positioning accuracy of the OBI isocenter and the CBCT image quality. For example, on one linac, the position of the marker on the cone-beam radiographic projections depends on the angular view and has maximal shifts of about 2 mm along the imager x-direction (patient's cross-plane). Sagging produces systematic shifts of the OBI isocenter as large as 1 mm posterior and 1 mm left in patient coordinates relative to the radiation isocenter. Further, it causes spatial distortion and blurring in CBCT image reconstructed from radiographic projections that are not corrected for OBI sagging. CBCT numbers vary by about 1% in full-fan scans and up to 3.5% in half-fan scans because of sagging. In order to achieve better localization accuracy in image-guided radiation therapy, sagging shifts of the kV OBI need to be corrected. In addition, correction of sagging improves image and provides better visualization of internal structures. Frequent quality assurance is required to monitor and maintain standards of variations in the mechanical accuracy of isocenter and image quality of CBCT because of sagging shifts.
引用
收藏
页码:180 / 194
页数:15
相关论文
共 26 条
[1]
ALHALABI A, 2008, INT J RADIAT ONCOL, V72, pS668
[2]
Measurement and Correction of Sagging Shifts in KV Cone-Beam CT From An On-Board Imaging System [J].
Ali, I. ;
Ahmad, S. .
MEDICAL PHYSICS, 2008, 35 (06)
[3]
3D image guided brachytherapy [J].
Devic, S. ;
Vuong, T. ;
Evans, M. ;
Podgorsak, E. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S547-S548
[4]
Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[5]
PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[6]
3D cone-beam CT reconstruction for circular trajectories [J].
Grass, M ;
Köhler, T ;
Proksa, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) :329-347
[7]
Image-guided radiotherapy for liver cancer using respiratorycorrelated computed tomography and cone-beam computed tomography [J].
Guckenberger, Matthias ;
Sweeney, Reinhart A. ;
Wilbert, Juergen ;
Krieger, Thomas ;
Richter, Anne ;
Baier, Kurt ;
Mueller, Gerd ;
Sauer, Otto ;
Flentje, Michael .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01) :297-304
[8]
Extracranial stereotactic radiation therapy:: Set-up accuracy of patients treated for liver metastases [J].
Herfarth, KK ;
Debus, J ;
Lohr, F ;
Bahner, ML ;
Fritz, P ;
Höss, A ;
Schlegel, W ;
Wannenmacher, MF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (02) :329-335
[9]
Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[10]
Cone-beam computed tomography with a flat-panel imager: Initial performance characterization [J].
Jaffray, DA ;
Siewerdsen, JH .
MEDICAL PHYSICS, 2000, 27 (06) :1311-1323