This paper compares the advantages and disadvantages of two different configurations for the extraction of triazines from water samples: (1) on-fibre solid-phase microextraction (SPME) coupled to conventional liquid chromatography (LC); and (2) in-tube SPME coupled to capillary LC. In-tube SPME has been effected either with a packed column or with an open capillary column. A critical evaluation of the main parameters affecting the performance of each method has been carried out in order to select the most suitable approach according to the requirements of the analysis. In the on-fibre SPME configuration the fibre coating was polydimethylsiloxane (PDMS)-divinylbenzene (DVB). The limits of detection (LODs) obtained with this approach under the optimized extraction and desorption conditions were between 25 and 125 mu g/l. The in-tube SPME approach with a C-18 packed column (35 turn x 0.5 turn I.D., 5 mu m particle size) connected to a switching micro-valve provided the best sensitivity; under such configuration the LODs were between 0.025 and 0.5 mu g/L. The in-tube SPME approach with an open capillary column coated with PDMS (30 cm x 0.25 mm I.D., 0.25 mu m of thickness coating) connected to the injection valve provided LODs between 0.1 and 0.5 mu g/L. In all configurations UV detection at 230 nm was used. Atrazine, simazine, propazine, ametryn, prometryn and terbutryn were selected as model compounds. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:159 / 171
页数:13
相关论文
共 30 条
[1]   Trace determination of priority pesticides in water by means of high-speed on-line solid-phase extraction-liquid chromatography-tandem mass spectrometry using turbulent-flow chromatography columns for enrichment and a short monolithic column for fast liquid chromatographic separation [J].
Asperger, A ;
Efer, J ;
Koal, T ;
Engewald, W .
JOURNAL OF CHROMATOGRAPHY A, 2002, 960 (1-2) :109-119
[2]   Evaluation of a field-test kit for triazine herbicides (Se nsioScreen® TR500) as a fast assay to detect pesticide contamination in water samples [J].
Ballesteros, B ;
Barceló, D ;
Dankwardt, A ;
Schneider, P ;
Marco, MP .
ANALYTICA CHIMICA ACTA, 2003, 475 (1-2) :105-115
[3]   Solid-phase microextraction in pesticide residue analysis [J].
Beltran, J ;
López, FJ ;
Hernández, F .
JOURNAL OF CHROMATOGRAPHY A, 2000, 885 (1-2) :389-404
[4]   Enzyme flow immunoassay using a Protein G column for the screening of triazine herbicides in surface and waste water [J].
Bjarnason, B ;
Chimuka, L ;
Önnerfjord, P ;
Eremin, SA ;
Jönsson, JÄ ;
Johansson, G ;
Emnéus, J .
ANALYTICA CHIMICA ACTA, 2001, 426 (02) :197-207
[5]   Gas chromatography/ion trap mass spectrometry applied for the anal sis of triazine herbicides in environmental waters by an isotope dilution technique [J].
Cai, ZW ;
Wang, DL ;
Ma, WT .
ANALYTICA CHIMICA ACTA, 2004, 503 (02) :263-270
[6]   Determination of triazines and dealkylated and hydroxylated metabolites in river water using a propazine-imprinted polymer [J].
Carabias-Martínez, R ;
Rodríguez-Gonzalo, E ;
Herrero-Hernández, E .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1085 (02) :199-206
[7]  
CRESCENZI C, 2000, ENCY ANAL CHEM, P6582
[8]   Determination of chlorotriazines, methylthiotriazines and one methoxytriazine by SPE-LC-UV in water samples [J].
Dopico, MS ;
González, MV ;
Castro, JM ;
González, E ;
Pérez, J ;
Rodríguez, M ;
Calleja, A ;
Vilariño, JML .
TALANTA, 2003, 59 (03) :561-569
[9]   Effect of the various parameters governing solid-phase microextraction for the trace-determination of pesticides in water [J].
Dugay, J ;
Miège, C ;
Hennion, MC .
JOURNAL OF CHROMATOGRAPHY A, 1998, 795 (01) :27-42
[10]  
Ellison S.L.R., 2000, GUIDE QUANTIFYING UN, V2nd ed.