Strategies to improve the signal and noise performance of active matrix, flat-panel imagers for diagnostic x-ray applications

被引:147
作者
Antonuk, LE [1 ]
Jee, KW
El-Mohri, Y
Maolinbay, M
Nassif, S
Rong, X
Zhao, Q
Siewerdsen, JH
Street, RA
Shah, KS
机构
[1] Univ Michigan, Med Ctr, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[2] Xerox Corp, Palo Alto Res Ctr, Palo Alto, CA 94304 USA
[3] Radiat Monitoring Devices Inc, Watertown, MA 02172 USA
关键词
active matrix flat-panel imager; cascaded systems analysis; diagnostic x-ray imaging; fluoroscopy; indirect detection; direct detection;
D O I
10.1118/1.598831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A theoretical investigation of factors limiting the detective quantum efficiency (DQE) of active matrix flat-panel imagers (AMFPIs), and of methods to overcome these limitations, is reported. At the higher exposure levels associated with radiography, the present generation of AMFPIs is capable of exhibiting DQE performance equivalent, or superior, to that of existing film-screen and computed radiography systems. However, at exposure levels commonly encountered in fluoroscopy, AMFPIs exhibit significantly reduced DQE and this problem is accentuated at higher spatial frequencies. The problem applies both to AMFPIs that rely on indirect detection as well as direct detection of the incident radiation. This reduced performance derives from the relatively large magnitude of the square of the total additive noise compared to the system gain for existing AMFPIs, In order to circumvent these restrictions, a variety of strategies to decrease additive noise and enhance system gain are proposed. Additive noise could be reduced through improved preamplifier, pixel and array design, including the incorporation of compensation lines to sample external line noise. System gain could be enhanced through the use of continuous photodiodes, pixel amplifiers, or higher gain x-ray converters such as lead iodide. The feasibility of these and other strategies is discussed and potential improvements to DQE performance are quantified through a theoretical investigation of a variety of hypothetical 200 mu m pitch designs. At low exposures, such improvements could greatly increase the magnitude of the low spatial frequency component of the DQE, rendering it practically independent of exposure while simultaneously reducing the falloff in DQE at higher spatial frequencies, Furthermore, such noise reduction and gain enhancement could lead to the development of AMFPIs with high DQE performance which are capable of providing both high resolution radiographic images, at similar to 100 mu m pixel resolution, as well as variable resolution fluoroscopic images at 30 fps. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01302-X].
引用
收藏
页码:289 / 306
页数:18
相关论文
共 44 条
  • [11] ENERGY AND ANGULAR-DEPENDENCE OF X-RAY ABSORPTION AND ITS EFFECT ON RADIOGRAPHIC RESPONSE IN SCREEN-FILM SYSTEMS
    CHAN, HP
    DOI, K
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (05) : 565 - 579
  • [12] A SPATIAL-FREQUENCY DEPENDENT QUANTUM ACCOUNTING DIAGRAM AND DETECTIVE QUANTUM EFFICIENCY MODEL OF SIGNAL AND NOISE-PROPAGATION IN CASCADED IMAGING-SYSTEMS
    CUNNINGHAM, IA
    WESTMORE, MS
    FENSTER, A
    [J]. MEDICAL PHYSICS, 1994, 21 (03) : 417 - 427
  • [13] Degradation of the detective quantum efficiency due to a non-unity detector fill factor
    Cunningham, IA
    [J]. PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 22 - 31
  • [14] X-RAY-IMAGING WITH AMORPHOUS SELENIUM - DETECTIVE QUANTUM EFFICIENCY OF PHOTOCONDUCTIVE RECEPTORS FOR DIGITAL MAMMOGRAPHY
    FAHRIG, R
    ROWLANDS, JA
    YAFFE, MJ
    [J]. MEDICAL PHYSICS, 1995, 22 (02) : 153 - 160
  • [15] HIGH-SENSITIVITY READOUT OF 2D A-SI IMAGE SENSORS
    FUJIEDA, I
    STREET, RA
    WEISFIELD, RL
    NELSON, S
    NYLEN, P
    PEREZMENDEZ, V
    CHO, GS
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1993, 32 (1A): : 198 - 204
  • [16] GHO G, 1993, IEEE T NUCL SCI, V40, P323
  • [17] Recent developments on a CdTe based X-ray detector for digital radiography
    Glasser, F
    Martin, JL
    Thevenin, B
    Schermesser, P
    Pantigny, P
    Laurent, JY
    Rambaud, P
    Pitault, B
    Paltrier, S
    [J]. PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 513 - 519
  • [18] Performance characteristics of an amorphous silicon flat panel X-ray imaging detector
    Granfors, PR
    [J]. MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 480 - 490
  • [19] JOHNS HE, 1983, PHYSICS RADIOLOGY, P558
  • [20] Dynamic X-ray imaging system based on an amorphous silicon thin-film array
    Jung, N
    Alving, PL
    Busse, F
    Conrads, N
    Meulenbrugge, HM
    Rutten, W
    Schiebel, U
    Weibrecht, M
    Wieczorek, H
    [J]. PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 396 - 407