High incidence of appropriate and inappropriate ICD therapies in children and adolescents with implantable cardioverter defibrillator

被引:100
作者
Korte, T
Köditz, H
Niehaus, M
Paul, T
Tebbenjohanns, J
机构
[1] Hannover Med Sch, Dept Cardiol & Pediat Cardiol, D-30625 Hannover, Germany
[2] Med Univ S Carolina, Dept Pediat Cardiol, Charleston, SC USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2004年 / 27卷 / 07期
关键词
implantable cardioverter defibrillator; pediatric cardiology; appropriate and inappropriate therapies;
D O I
10.1111/j.1540-8159.2004.00560.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Appropriate and inappropriate therapies of implantable cardioverter defibrillators have a major impact on morbidity and quality of life in ICD recipients, but have not been systematically studied in children and young adults during long-term follow-up. ICD implantation was performed in 20 patients at the mean age of 16 +/- 6 years, 11 of which had prior surgical repair of a congenital heart defect, 9 patients had other cardiac diseases. Implant indications were aborted sudden cardiac death in six patients, recurrent ventricular tachycardia in 9 patient, and syncope in 5 patients. Epicardial implantation was performed in 6 and transvenous implantation in 14 patients. Incidence, reasons and predictors (age, gender, repaired congenital heart disease, history of supraventricular tachycardia, and epicardial electrode system) of appropriate and inappropriate ICD therapies were analyzed during a mean follow-up period of 51 +/- 31 months range 18-132 months. There were a total 239 ICD therapies in 17 patients (85%) with a therapy rate of 2.8 per patient-years of follow-up. 127 (53%)ICD therapies in 15 (75%) patients were catagorized as appropriate and 112 (47%) therapies in 10 (50%) patients us inappropriate, with a rate of 1.5 appropriate and 1.3 inappropriate ICD therapies per patient-years of follow-up. Time to first appropriate therapy was 16 +/- 18 months. Appropriate therapies were caused by ventricular fibrillation in 29 and ventricular tachycardia in 98 episodes. Termination was successful by antitachycardia pacing in 4 (3%) and by shock therapy in 123 episodes (97%). Time to first inappropriate therapy was 16 +/- 17 months. Inappropriate therapies were caused by supraventricular tachycardia in 77 (69%), T wave oversensing in 19 (17%), and electrode defect in 16 episodes (14%). It caused shocks in 87 (78%) and only antitachycardia pacing in 25 episodes (22%). No clinical variable could be identified as predictor of either appropriate or inappropriate ICD therapies. There is a high rate of ICD therapies in young ICD recipients, the majority of which occur during early follow-up. The rate of inappropriate therapies is as high as 47% and is caused by supraventricular tachycardia and electrode complications in the majority of cases. Prospective trials ore required to establish preventative startegies of ICD therapies in this young patient population.
引用
收藏
页码:924 / 932
页数:9
相关论文
共 37 条
[11]   Prospective evaluation of catheter ablation in patients with implantable cardioverter defibrillators and multiple inappropriate ICD therapies due to atrial fibrillation and type I atrial flutter [J].
Korte, T ;
Niehaus, M ;
Meyer, O ;
Tebbenjohanns, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (07) :1061-1066
[12]   Hospital readmission after transvenous cardioverter/defibrillator implantation -: A single centre study [J].
Korte, T ;
Jung, W ;
Ostermann, G ;
Wolpert, C ;
Spehl, S ;
Esmailzadeh, B ;
Lüderitz, B .
EUROPEAN HEART JOURNAL, 2000, 21 (14) :1186-1191
[13]   PRELIMINARY EXPERIENCE WITH NONTHORACOTOMY IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS IN YOUNG-PATIENTS [J].
KRON, J ;
SILKA, MJ ;
OHM, OJ ;
BARDY, G ;
BENDITT, D ;
CHILSON, D ;
GILLIS, AM ;
MITCHELL, LB ;
GULAMHUSEIN, S ;
KUCK, K ;
SIEBELS, J ;
BREITHARDT, G ;
BLOCK, M ;
TOUBOUL, P ;
PRADEL, L ;
BRACHMANN, J ;
RUSKIN, J ;
HAFFAJEE, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01) :26-30
[14]   Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing [J].
Kühlkamp, V ;
Dörnberger, V ;
Mewis, C ;
Suchalla, R ;
Bosch, RF ;
Seipel, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (07) :905-915
[15]   Preliminary clinical experience with the first dual chamber pacemaker defibrillator [J].
Lavergne, T ;
Daubert, JC ;
Chauvin, M ;
Dolla, E ;
Kacet, S ;
Leenhardt, A ;
Mabo, P ;
Ritter, P ;
Sadoul, N ;
Saoudi, N ;
Henry, C ;
Nitzsche, R ;
Ripart, A ;
Murgatroyd, F .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01) :182-188
[16]   DIFFERENTIAL EFFECT OF INTRAVENOUS PROCAINAMIDE ON ANTEROGRADE AND RETROGRADE ACCESSORY PATHWAY REFRACTORINESS [J].
LEITCH, JW ;
KLEIN, GJ ;
YEE, R ;
FELDMAN, RD ;
BROWN, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (01) :118-124
[17]   Supraventricular arrhythmias in children and young adults with implantable cardioverter defibrillators [J].
Love, BA ;
Barrett, KS ;
Alexander, ME ;
Bevilacqua, LM ;
Epstein, MR ;
Triedman, JK ;
Walsh, EP ;
Berul, CI .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (10) :1097-1101
[18]   PATIENT ACCEPTANCE OF THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR IN VENTRICULAR TACHYARRHYTHMIAS [J].
LUDERITZ, B ;
JUNG, W ;
DEISTER, A ;
MARNEROS, A ;
MANZ, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (09) :1815-1821
[19]  
McAnulty J, 1997, NEW ENGL J MED, V337, P1576
[20]   Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia [J].
Moss, AJ ;
Hall, WJ ;
Cannom, DS ;
Daubert, JP ;
Higgins, SL ;
Klein, H ;
Levine, JH ;
Saksena, S ;
Waldo, AL ;
Wilber, D ;
Brown, MW ;
Heo, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (26) :1933-1940