Study on the Infrared Lens-Free Irradiation Thermometer Based on InGaAs Detector at NIM

被引:4
作者
Hao, X. P. [1 ]
Yuan, Z. D. [1 ]
Huang, S. Y. [1 ,2 ]
Song, J. [1 ]
Xu, K. [1 ,2 ]
机构
[1] Natl Inst Metrol, Div Thermophys & Proc Measurements, Beijing 100013, Peoples R China
[2] Chengdu Univ Technol, Coll Appl Nucl Technol & Automat Engn, Chengdu 610059, Peoples R China
基金
中国国家自然科学基金;
关键词
Heat-pipe blackbody; Lens-free irradiation thermometer; Multipoint calibration method; Radiation thermometry;
D O I
10.1007/s10765-015-1955-7
中图分类号
O414.1 [热力学];
学科分类号
摘要
A lens-free irradiation thermometer (LF-IRRT) designed for laboratory applications, with center wavelength based on the InGaAs detector, and a temperature range from to , has been developed at NIM of China. The thermometer has no imaging optics for collecting the radiation, for which a major benefit is improved stability compared to other radiation thermometers. This paper introduces the measurement principle, the structure of the LF-IRRT, the multiple-point calibration method, and the stability and expanded uncertainty of the thermometer. The stability of the LF-IRRT is better than at temperatures from to over a period of 1 month. The expanded uncertainty is less than for temperatures between and . The size-of-source effect of the LF-IRRT is 0.6 % from the source diameter range from 10 mm to 65 mm.
引用
收藏
页码:3320 / 3329
页数:10
相关论文
共 12 条
[1]  
[Anonymous], INT J THERMOPH UNPUB
[2]  
[Anonymous], CHIN J SCI INSTRUM
[3]  
Gutschwager B., 2004, P TEMPMEKO 2004 9 IN, P605
[4]   Size-of-Source Effect Difference Between Direct and Indirect Methods of Radiation Thermometers [J].
Hao, X. ;
Yuan, Z. ;
Lu, X. ;
Zhao, W. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (7-8) :1655-1663
[5]  
Hao X.P., 2010, CHIN J SCI INSTRUM, V31, P223
[6]   A comparison of the In, Sn, Zn and Al fixed points by radiation thermometry between NIM and NPL and verification of the NPL blackbody reference sources from 156 °C to 1000 °C [J].
Hao, Xiaopeng ;
McEvoy, Helen ;
Machin, Graham ;
Yuan, Zundong ;
Wang, Tiejun .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2013, 24 (07)
[7]  
Lohrengel J, 1996, PTB-MITT, V106, P259
[8]  
Sakuma F., 1982, TEMPERATURE ITS MEAS, V5, P421
[9]  
Sapritsky S P, 1992, METROLOGIA, V29, P9
[10]   Uncertainty budgets for calibration of radiation thermometers below the silver point [J].
Saunders, P. ;
Fischer, J. ;
Sadli, M. ;
Battuello, M. ;
Park, C. W. ;
Yuan, Z. ;
Yoon, H. ;
Li, W. ;
van der Ham, E. ;
Sakuma, F. ;
Ishii, J. ;
Ballico, M. ;
Machin, G. ;
Fox, N. ;
Hollandt, J. ;
Matveyev, M. ;
Bloembergen, P. ;
Ugur, S. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (03) :1066-1083