Biomechanics of lateral lumbar interbody fusion constructs with lateral and posterior plate fixation

被引:61
作者
Fogel, Guy R.
Parikh, Rachit D. [1 ]
Ryu, Stephen I. [2 ,3 ]
Turner, Alexander W. L. [1 ]
机构
[1] NuVasive Inc, San Diego, CA USA
[2] Palo Alto Med Fdn, Palo Alto, CA USA
[3] Stanford Univ, Stanford, CA 94305 USA
关键词
lumbar interbody fusion; XLIF; stability; fixation; range of motion; PEDICLE SCREW PLACEMENT; BACK MUSCLE INJURY; INTERSPINAL PROCESS DECOMPRESSION; MINIMALLY INVASIVE SURGERY; SPINE SURGERY; HISTOCHEMICAL ANALYSES; INVITED SUBMISSION; PERIPHERAL-NERVES; CLINICAL ARTICLE; MARCH; 2005;
D O I
10.3171/2013.11.SPINE13617
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Object. Lumbar interbody fusion is indicated in the treatment of degenerative conditions. Laterally inserted interbody cages significantly decrease range of motion (ROM) compared with other cages. Supplemental fixation options such as lateral plates or spinous process plates have been shown to provide stability and to reduce morbidity. The authors of the current study investigate the in vitro stability of the interbody cage with a combination of lateral and spinous process plate fixation and compare this method to the established bilateral pedicle screw fixation technique. Methods. Ten L1-5 specirnens were evaluated using multidirectional nondestructive moments (+/- 7.5 Nm), with a custom 6 degrees-of-freedom spine simulator. Intervertebral motions (ROM) were measured optoelectronically. Each spine was evaluated under the following conditions at the L3-4 level: intact; interbody cage alone (stand-alone); cage supplemented with lateral plate; cage supplemented with ipsilateral pedicle screws; cage supplemented with bilateral pedicle screws; cage supplemented with spinous process plate; and cage supplemented with a combination of lateral plate and spinous process plate. Intervertebral rotations were calculated, and ROM data were normalized to the intact ROM data. Results. The stand-alone laterally inserted interbody cage significantly reduced ROM with respect to the intact state in flexion-extension (31.6% intact ROM, p < 0.001), lateral bending (32.5%, p < 0.001), and axial rotation (69.4%, p = 0.002). Compared with the stand-alone condition, addition of a lateral plate to the interbody cage did not significantly alter the ROM in flexion-extension (p = 0.904); however, it was significantly decreased in lateral bending and axial rotation (p < 0.001). The cage supplemented with a lateral plate was not statistically different from bilateral pedicle screws in lateral bending (p = 0.579). Supplemental fixation using a spinous process plate was not significantly different from bilateral pedicle screws in flexion-extension (p = 0.476). The combination of lateral plate and spinous process plate was not statistically different from the cage supplemented with bilateral pedicle screws in all the loading modes (p >= 0.365). Conclusions. A combination of lateral and spinous process plate fixation to supplement a laterally inserted interbody cage helps achieve rigidity in all motion planes similar to that achieved with bilateral pedicle screws.
引用
收藏
页码:291 / 297
页数:7
相关论文
共 37 条
[1]
Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patients Clinical article [J].
Amato, Vincenzo ;
Giannachi, Luigi ;
Irace, Claudio ;
Corona, Claudio .
JOURNAL OF NEUROSURGERY-SPINE, 2010, 12 (03) :306-313
[2]
ANALYSIS OF COMPLICATIONS IN PATIENTS TREATED WITH THE X-STOP INTERSPINOUS PROCESS DECOMPRESSION SYSTEM: PROPOSAL FOR A NOVEL ANATOMIC SCORING SYSTEM FOR PATIENT SELECTION AND REVIEW OF THE LITERATURE [J].
Barbagallo, Giuseppe M. V. ;
Olindo, Giuseppe ;
Corbino, Leonardo ;
Albanese, Vincenzo .
NEUROSURGERY, 2009, 65 (01) :111-120
[3]
Bess R, 2008, EXTREME LATERAL INTE, P31
[4]
Biomechanical Analysis and Review of Lateral Lumbar Fusion Constructs [J].
Cappuccino, Andrew ;
Cornwall, G. Bryan ;
Turner, Alexander W. L. ;
Fogel, Guy R. ;
Duong, Huy T. ;
Kim, Kee D. ;
Brodke, Darrel S. .
SPINE, 2010, 35 (26) :S361-S367
[5]
Dooley Zachary A., 2013, International Journal of Experimental and Computational Biomechanics, V2, P189
[6]
FUSIONS AND TRANSFUSIONS - AN ANALYSIS OF BLOOD-LOSS AND AUTOLOGOUS REPLACEMENT DURING LUMBAR FUSIONS [J].
JOHNSON, RG ;
MURPHY, M ;
MILLER, M .
SPINE, 1989, 14 (04) :358-362
[7]
Biomechanics of a Lumbar Interspinous Anchor with Transforaminal Lumbar Interbody Fixation [J].
Kaibara, Taro ;
Karahalios, Dean G. ;
Porter, Randall W. ;
Kakarla, Udaya K. ;
Reyes, Phillip M. ;
Choi, Seok Kwang ;
Yaqoobi, Ali S. ;
Crawford, Neil R. .
WORLD NEUROSURGERY, 2010, 73 (05) :572-577
[8]
Biomechanics of a lumbar interspinous anchor with anterior lumbar interbody fusion Laboratory investigation [J].
Karahalios, Dean G. ;
Kaibara, Taro ;
Porter, Randall W. ;
Kakarla, Udaya K. ;
Reyes, Phillip M. ;
Baaj, Ali A. ;
Yaqoobi, Ali S. ;
Crawford, Neil R. .
JOURNAL OF NEUROSURGERY-SPINE, 2010, 12 (04) :372-380
[9]
BACK MUSCLE INJURY AFTER POSTERIOR LUMBAR SPINE SURGERY .1. HISTOLOGIC AND HISTOCHEMICAL ANALYSES IN RATS [J].
KAWAGUCHI, Y ;
MATSUI, H ;
TSUJI, H .
SPINE, 1994, 19 (22) :2590-2597
[10]
BACK MUSCLE INJURY AFTER POSTERIOR LUMBAR SPINE SURGERY .2. HISTOLOGIC AND HISTOCHEMICAL ANALYSES IN HUMANS [J].
KAWAGUCHI, Y ;
MATSUI, H ;
TSUJI, H .
SPINE, 1994, 19 (22) :2598-2602