Optimization of a comprehensive two-dimensional normal-phase and reversed-phase liquid chromatography system

被引:25
作者
Dugo, Paola
Fernandez, Maria del Mar Ramirez
Cotroneo, Antonella
Dugo, Giovanni
Mondello, Luigi
机构
[1] Univ Messina, Fac Sci, Dipartimento Chim Organ & Biol, I-98166 Messina, Italy
[2] Natl Inst Criminalist & Criminol, B-1120 Brussels, Belgium
[3] Univ Messina, Fac Farm, Dipartimento Farmacochim, I-98168 Messina, Italy
关键词
D O I
10.1093/chromsci/44.9.561
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
The present investigation is based on the evaluation of the performance of a comprehensive two-dimensional liquid chromatography (LCxLC) system during method optimization. The LCxLC set-up, operated in normal phase (NP) mode (adsorption) in the first dimension (11 D) and reversed-phase (RP) mode in the second dimension (2D), is equipped with a 1D microbore silica column and a 2D monolithic C-18 column with a 10-port two position valve as the interface. A photodiode array detector is used after the 2D separation. A possible cause of peak distorsion because of the immiscibility of the mobile phases employed in the two dimensions is resolved. The optimization of the analytical run time and flow rate for both dimensions and the initial gradient in the 2D is carried out with various standard compounds. The potential and versatility of this LCxLC approach is demonstrated through the separation of 11 standard components, most of them allergens. The latter, which are characterized by a scattered distribution on the 2D space plane, underwent separation on both a hydrophobicity and polarity basis.
引用
收藏
页码:561 / 565
页数:5
相关论文
共 20 条
[1]
DEVELOPMENTS IN MULTIDIMENSIONAL SEPARATION SYSTEMS [J].
CORTES, HJ .
JOURNAL OF CHROMATOGRAPHY, 1992, 626 (01) :3-23
[2]
Quantitation of suspected allergens in fragrances Part II. Evaluation of comprehensive gas chromatography-conventional mass spectrometry [J].
Debonneville, C ;
Chaintreau, A .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1027 (1-2) :109-115
[3]
Comprehensive two-dimensional normal-phase (adsorption) - Reversed-phase liquid chromatography [J].
Dugo, P ;
Favoino, O ;
Luppino, R ;
Dugo, G ;
Mondello, L .
ANALYTICAL CHEMISTRY, 2004, 76 (09) :2525-2530
[4]
GIDDINGS JC, 1990, MULTIDIMENSIONAL CHR, P1
[5]
INJECTION OF ELUITES IN SOLVENTS STRONGER THAN THE MOBILE PHASE IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
HOFFMAN, NE ;
PAN, SL ;
RUSTUM, AM .
JOURNAL OF CHROMATOGRAPHY, 1989, 465 (02) :189-200
[6]
Phase system selectivity and two-dimensional separations in liquid column chromatography [J].
Jandera, P ;
Halama, M ;
Kolárová, L ;
Fischer, J ;
Novotná, K .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1087 (1-2) :112-123
[7]
Lubda D., 2001, LC GC EUR, V14, P730
[8]
MULTIDIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
MAJORS, RE .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1980, 18 (10) :571-579
[9]
Assessment of injection volume limits when using on-column focusing with microbore liquid chromatography [J].
Mills, MJ ;
Maltas, J ;
Lough, WJ .
JOURNAL OF CHROMATOGRAPHY A, 1997, 759 (1-2) :1-11
[10]
Silver-ion reversed-phase comprehensive two-dimensional liquid chromatography combined with mass spectrometric detection in lipidic food analysis [J].
Mondello, L ;
Tranchida, PQ ;
Stanek, V ;
Jandera, P ;
Dugo, G ;
Dugo, P .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1086 (1-2) :91-98