Automatic exposure control to reduce the dose in subsecond multislice spiral CT: Phantom measurements and clinical results

被引:19
作者
Greess, H
Wolf, H
Suess, C
Kalender, WA
Bautz, W
Baum, U
机构
[1] Univ Erlangen Nurnberg, Inst Diagnost Radiol, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Inst Med Phys, Erlangen, Germany
[3] Siemens AG Med Solut, Unternehmensbereich, Forchheim, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2004年 / 176卷 / 06期
关键词
multislice spiral CT; dosimetry; automatic exposure control; dose reduction; tube current modulation;
D O I
10.1055/s-2004-813183
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the potential of dose reduction in multislice spiral CT (MSCT) with automatic exposure control. Materials and Methods: The study was performed on a Sensation 4 multislice scanner. This prototype implementation analyzed the distribution of the attenuation along the z-axis in the lateral and sagittal directions of the digital radiogram. Depending on this distribution of the attenuation, the tube current (mA) is defined for every tube rotation. In addition, the tube current was modulated during each tube rotation. First, a three step oval water phantom was measured to evaluate the potential of this method with respect to dose reduction and image quality. In a patient study (n = 26), four different scan regions (shoulder, thorax, abdomen, pelvis) were examined and dose (mAs) and image quality evaluated in comparison to examinations with a standard protocol for these regions in adults and a weight-adjusted standard protocol in children. The image quality was classified in consensus as good, sufficient and poor image quality. Results: By adapting and modulating the tube current, we substantially reduced the variation of noise in one spiral scan and in one scan region of our patient collective. The dose (average mAs) was reduced by 31% to 66% in children (mean 44%) and between 35% and 64% in adults (mean 53%), depending on the scan region. The image quality was substantially improved in regions with high attenuation and did not suffer in low attenuation regions. Conclusion: The dose can be reduced substantially by an automatic exposure control including angular tube current modulation with the same or improved image quality.
引用
收藏
页码:862 / 869
页数:8
相关论文
共 17 条
[1]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[2]   Minimizing radiation dose for pediatric body applications of single-detector helical CT: Strategies at a large children's hospital [J].
Donnelly, LF ;
Emery, KH ;
Brody, AS ;
Laor, T ;
Gylys-Morin, VM ;
Anton, CG ;
Thomas, SR ;
Frush, DP .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :303-306
[3]  
Fox S, 2002, MED PHYS, V29, P1359
[4]  
Giacomuzzi S M, 1996, Aktuelle Radiol, V6, P110
[5]   Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose) [J].
Greess, H ;
Nömayr, A ;
Wolf, H ;
Baum, U ;
Lell, M ;
Böwing, B ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2002, 12 (06) :1571-1576
[6]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[7]   Dose reduction in spiral CT:: Detection of pulmonary nodules with and without anatomic adaptation of tube current. [J].
Greess, H ;
Baum, U ;
Wolf, H ;
Lell, M ;
Nömayr, A ;
Schmidt, B ;
Kalender, WA ;
Bautz, W .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2001, 173 (05) :466-470
[8]  
Greess H, 1999, ROFO-FORTSCHR RONTG, V170, P246
[9]  
Herzog P, 2002, RADIOLOGE, V42, P691, DOI 10.1007/s00117-002-0799-x
[10]   Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers [J].
Kalender, WA ;
Wolf, H ;
Suess, C ;
Gies, M ;
Greess, H ;
Bautz, WA .
EUROPEAN RADIOLOGY, 1999, 9 (02) :323-328