Evaluation and quality assessment of glucose concentration measurement in blood by point-of-care testing devices

被引:21
作者
Wehmeier, Michael [1 ]
Arndt, Burkhard T. [1 ]
Schumann, Gerhard [1 ]
Kulpmann, Wolf R. [1 ]
机构
[1] Hannover Med Sch, Dept Clin Chem, D-30625 Hannover, Germany
关键词
comparative method; glucose; metrological; traceability; point-of-care testing (POCT); RECOMMENDATION;
D O I
10.1515/CCLM.2006.141
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
100118 [医学信息学]; 100208 [临床检验诊断学];
摘要
Background: In this study the analytical performance of eight glucose point-of-care testing (POCT) devices was evaluated. For this purpose, POCT measurement of glucose in heparinized blood collected from patients was paralleled by determination of the glucose concentration in the respective plasma by an analyzer (Hitachi 917) in the central laboratory, providing traceable results. Methods: Trueness of POCT measurements was studied by comparing the plasma POCT values ( mean of five measurements) with the results from the traceable measurement procedure (TMP). Results: The percentage of POCT results within +/- 6% of the TMP mean value ranged from 24% to 50%, depending on the POCT device. Within the reference interval of plasma glucose (4.4 - 6.0 mmol/L), up to 67% of the POCT values were lower than 4.4 mmol/L, leading to a false diagnosis of hypoglycemia. In the hypoglycemic range (-4.4 mmol/L) up to 29% of the POCT analyses were false normoglycemic. Conclusions: In conclusion, this study shows an insufficient trueness of glucose measurements by POCT devices in the normo- and hypoglycemic range. To improve quality assessment, sample splitting and simultaneous measurement of blood glucose concentration every 4 weeks by POCT devices and of plasma glucose concentration by a reliable TMP is recommended.
引用
收藏
页码:888 / 893
页数:6
相关论文
共 10 条
[1]
Clark N, 2003, DIABETES CARE, V26, P3333
[2]
Approved IFCC recommendation on reporting results for blood glucose (abbreviated) [J].
D'Orazio, P ;
Burnett, RW ;
Fogh-Andersen, N ;
Jacobs, E ;
Kuwa, K ;
Külpmann, WR ;
Larsson, L ;
Lewenstam, A ;
Maas, AHJ ;
Mager, G ;
Naskalski, JW ;
Okorodudu, AO .
CLINICAL CHEMISTRY, 2005, 51 (09) :1573-1576
[3]
Fogh-Andersen N, 1998, CLIN CHEM, V44, P655
[4]
DIRECT READING GLUCOSE ELECTRODES DETECT THE MOLALITY OF GLUCOSE IN PLASMA AND WHOLE-BLOOD [J].
FOGHANDERSEN, N ;
WIMBERLEY, PD ;
THODE, J ;
SIGGAARDANDERSEN, O .
CLINICA CHIMICA ACTA, 1990, 189 (01) :33-38
[5]
Hannestad U, 1997, CLIN CHEM, V43, P794
[6]
Evaluation of ''point of care'' devices in the measurement of low blood glucose in neonatal practice [J].
Ho, HT ;
Yeung, WKY ;
Young, BWY .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2004, 89 (04) :F356-F359
[7]
Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care hospital ambulatory patients in the United States and Canada [J].
Kost, GJ ;
Vu, HT ;
Lee, JH ;
Bourgeois, P ;
Kiechle, FL ;
Martin, C ;
Miller, SS ;
Okorodudu, AO ;
Podczasy, JJ ;
Webster, R ;
Whitlow, KJ .
CRITICAL CARE MEDICINE, 1998, 26 (03) :581-590
[8]
USE OF ARTERIAL BLOOD WITH BEDSIDE GLUCOSE REFLECTANCE METERS IN AN INTENSIVE-CARE UNIT - ARE THEY ACCURATE [J].
MASER, RE ;
BUTLER, MA ;
DECHERNEY, GS .
CRITICAL CARE MEDICINE, 1994, 22 (04) :595-599
[9]
*RICHTL BUND QUAL, 2001, DT ARZTEBLATT, V98, pA2747
[10]
SIGGAARDANDERSEN O, 1987, J CLIN CHEM CLIN BIO, V25, P369