Miniaturized capillary electrophoresis system with ultraviolet photometric detection combined with flow injection sample introduction using a modified falling-drop interface

被引:23
作者
Cao, XD [1 ]
Fang, Q [1 ]
Fang, ZL [1 ]
机构
[1] Zhejiang Univ, Inst Microanalyt Syst, Dept Chem, Hangzhou 310028, Peoples R China
基金
中国国家自然科学基金;
关键词
miniaturization; capillary electrophoresis; falling-drop sample introduction interface; flow-injection; UV-Vis detection;
D O I
10.1016/j.aca.2004.03.028
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A miniaturized capillary electrophoresis (CE) system with UV-Vis detection was coupled to a flow injection (FI) system for achieving high throughput continuous sample introduction. The cassette of a commercial CE instrument was modified to hold a 6.5 cm long silica capillary and a flow-through waste reservoir. The cassette was inserted into the flow-cell chamber of a commercial UV detector, with the light beam focused on the capillary and collected by two ball lenses on the cassette. The capillary inlet, left outside the cassette and detector, was positioned on the top of a vertical 3.5 mm. diameter glass rod, in close contact with an electrode. Samples injected through the FI system dropped freely on top of the pillar, covering the capillary inlet and electrode. Continuous sample introduction was achieved for CE separations under non-interrupted separation voltage, which was isolated from the FI system through the discontinuity of droplets. The newly developed interface and UV detection system was used for fast separation of sulphamethoxazole (SMZ) and trimethoprim (TMP) in sulphatrim tablets, achieving a high throughput of over 48 h(-1), and a low carryover of 2%. Separation efficiencies of 8 mum plate height and detection limits of 1.0mg 1(-1) for SMZ and 0.5 mg 1(-1) (3sigma) for TMP were obtained. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:473 / 479
页数:7
相关论文
共 19 条
[1]  
CAO XD, 2002, P 1 NAT S U TAS 2002, P46
[2]   Combination of flow injection with capillary electrophoresis. Part 5. Automated preconcentration and determination of pseudoephedrine in human plasma [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 394 (01) :13-22
[3]   Combination of flow injection with capillary electrophoresis - Part 4. Automated multicomponent monitoring of drug dissolution [J].
Chen, HW ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1998, 376 (02) :209-220
[4]   A high-throughput continuous sample introduction interface for microfluidic chip-based capillary electrophoresis systems [J].
Fang, Q ;
Xu, GM ;
Fang, ZL .
ANALYTICAL CHEMISTRY, 2002, 74 (06) :1223-1231
[5]  
FANG Q, 1999, ANAL CHIM ACTA, V27, P390
[6]   Combination of flow injection with capillary electrophoresis .1. The basic system [J].
Fang, ZL ;
Liu, ZS ;
Shen, Q .
ANALYTICA CHIMICA ACTA, 1997, 346 (02) :135-143
[7]   Combination of flow injection with capillary electrophoresis - Part 7. Microchip capillary electrophoresis system with flow injection sample introduction and amperometric detection [J].
Fu, CG ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 2000, 422 (01) :71-79
[8]   Combination of flow injection with capillary electrophoresis 8. Miniaturized capillary electrophoresis system with flow injection sample introduction and electrogenerated chemiluminescence detection [J].
Huang, XJ ;
Wang, SL ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 2002, 456 (02) :167-175
[9]   Capillary electrophoresis system with flow injection sample introduction and chemiluminescence detection on a chip platform [J].
Huang, XJ ;
Pu, QS ;
Fang, ZL .
ANALYST, 2001, 126 (03) :281-284
[10]   On-line coupling of gas diffusion to capillary electrophoresis [J].
Kuban, P ;
Karlberg, B .
TALANTA, 1998, 45 (03) :477-484