Bone and Muscle Parameters of the Tibia: Agreement Between the XCT 2000 and XCT 3000 Instruments

被引:16
作者
Burrows, Melonie [1 ,2 ]
Cooper, David M. L. [3 ]
Liu, Danmei [1 ]
McKay, Heather A. [1 ]
机构
[1] Univ British Columbia, Dept Orthopaed, Vancouver, BC V5Z 1L8, Canada
[2] Univ British Columbia, Fac Med, Dept Orthopaed & Engn, Ctr Hip Hlth & Mobil,Vancouver Coastal Hlth Res I, Vancouver, BC V5Z 1L8, Canada
[3] Univ Saskatchewan, Dept Anat & Cell Biol, Saskatoon, SK S7N 0W0, Canada
关键词
Bone geometry; bone strength; peripheral quantitative computed tomography; tibia; QUANTITATIVE COMPUTED-TOMOGRAPHY; CROSS-CALIBRATION; 20-MONTH PQCT; IN-VITRO; ACCURACY; PRECISION; STRENGTH; GEOMETRY; DENSITY; RADIUS;
D O I
10.1016/j.jocd.2008.09.005
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Peripheral quantitative computed tomography is a valuable tool to assess bone in children across growth, with long-term studies capturing nuances missed in cross-sectional studies. As children grow, a change from XCT 2000 to a XCT 3000 may be required to accommodate the increasing size of the lower limbs. We examined the precision and agreement between the Stratec XCT 2000 and 3000 on selected bone and muscle parameters. Twenty-eight participants (mean +/- SD; age 27.5 +/- 6.5 yr) underwent scans at the distal (8%), mid (50%), and proximal (66%) tibia sites, to assess total bone area, total bone density, and trabecular density (8% site); and total bone area, cortical area, cortical density (CoD), polar strength-strain index, and muscle cross-sectional area (50% and 66% sites). Outcomes between instruments were highly correlated; r = 0.90-0.99 for CoD across sites, with r = 0.97-0.99 for all other measures. Bland and Altman plots showed excellent agreement between instruments for all variables. Regression indicated no significant relationship between instrument and size of measurement (p > 0.05). Coefficients of variation were lower than previously reported (0.4-2.4%). For longitudinal studies, the XCT 3000 can replace the XCT 2000 with minimal influence on bone and muscle parameters.
引用
收藏
页码:186 / 194
页数:9
相关论文
共 27 条
[1]   Accuracy of cortical and trabecular bone measurements with peripheral quantitative computed tomography (pQCT) [J].
Augat, P ;
Gordon, CL ;
Lang, TF ;
Iida, H ;
Genant, HK .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2873-2883
[2]   Comparison and cross-calibration of DXA systems: ODX-240 and sophos L-XRA versus hologic QDR-4500, for spinal bone mineral measurement. Translation of a reference database [J].
Benmalek, A ;
Sabatier, JP .
OSTEOPOROSIS INTERNATIONAL, 1998, 8 (06) :570-577
[3]   pQCT measurement of bone parameters in young children - Validation of technique [J].
Binkley, TL ;
Specker, BL .
JOURNAL OF CLINICAL DENSITOMETRY, 2000, 3 (01) :9-14
[4]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Clinical evaluation of a high-resolution new peripheral quantitative computerized tomography (pQCT) scanner for the bone densitometry at the lower limbs [J].
Braun, MJ ;
Meta, MD ;
Schneider, P ;
Reiners, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2279-2294
[7]   FOREARM BMD AS MEASURED BY PERIPHERAL QUANTITATIVE COMPUTED-TOMOGRAPHY (PQCT) IN A GERMAN REFERENCE POPULATION [J].
BUTZ, S ;
WUSTER, C ;
SCHEIDTNAVE, C ;
GOTZ, M ;
ZIEGLER, R .
OSTEOPOROSIS INTERNATIONAL, 1994, 4 (04) :179-184
[8]   CROSS-CALIBRATION OF DXA EQUIPMENT - UPGRADING FROM A HOLOGIC QDR 1000/W TO A QDR-2000 [J].
FAULKNER, KG ;
GLUER, CC ;
ESTILO, M ;
GENANT, HK .
CALCIFIED TISSUE INTERNATIONAL, 1993, 52 (02) :79-84
[9]  
Genant HK, 1996, J BONE MINER RES, V11, P707
[10]  
GRAMPP S, 1995, J BONE MINER RES, V10, P1566