Elapsed time for capsular apposition to intraocular lens after cataract surgery

被引:48
作者
Hayashi, H
Hayashi, K
Nakao, F
Hayashi, F
机构
[1] Hayashi Eye Hosp, Hakata Ku, Fukuoka 812, Japan
[2] Fukuoka Univ, Sch Med, Dept Ophthalmol, Fukuoka 81401, Japan
关键词
D O I
10.1016/S0161-6420(02)01112-0
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To examine when the anterior and posterior lens capsule completely become apposed to optics of silicone and acrylic intraocular lenses (IOLs) implanted after cataract surgery and to determine whether the different IOL materials influence the timing of completion of capsular contact. Design: Randomized controlled clinical trial. Participants: Seventy eyes of 70 patients who were scheduled to undergo cataract surgery were randomly assigned to two groups using random number tables based on the type of IOL implanted: silicone or acrylic. Thirty-two patients in each group completed the follow-up. Intervention: All eyes underwent phacoemulsification surgery with implantation of either a silicone or acrylic IOL. All IOLs were accurately placed into the capsular bag. Main Outcome Measures: Contact of the anterior and posterior lens capsule with the IOL optic surface was evaluated using the Scheimpflug videophotography system at 3, 5, 7, 9, 11, 14, 21, and 28 days after surgery. The postoperative day at which each capsule was completely apposed to the IOL optic was determined. In addition, anterior chamber depth was also measured. Results: The anterior capsule was in contact with the IOL optic on the same day or earlier than the posterior capsule in all patients. Complete apposition of the IOL was observed significantly earlier with silicone IOLs than with acrylic IOLs with both the anterior capsule (6.2 versus 3.6 postoperative days; P < 0.0001) and the posterior capsule (11.1 versus 7.4 postoperative days; P = 0.0339). No significant change in mean anterior chamber depth was observed with the silicone IOL, whereas there was significant anterior shift after implantation of the acrylic IOL. Conclusions: Capsular contact with the IOL optic is completed within approximately 8 days after cataract surgery with silicone IOLs and 11 days with acrylic IOLs. Complete apposition to both the anterior and posterior capsule was significantly earlier with silicone IOLs than with acrylic IOLs. (C) 2002 by the American Academy of Ophthalmology.
引用
收藏
页码:1427 / 1431
页数:5
相关论文
共 10 条
[1]  
Apple DJ, 2000, SURV OPHTHALMOL, V45, pS5
[2]   Accurate determination of effective lens position and lens-capsule distance with 4 intraocular lenses [J].
Findl, O ;
Drexler, W ;
Menapace, R ;
Bobr, B ;
Bittermann, S ;
Vass, C ;
Rainer, G ;
Hitzenberger, CK ;
Fercher, AF .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (08) :1094-1098
[3]   Reduction in the area of the anterior capsule opening after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation [J].
Hayashi, K ;
Hayashi, H ;
Nakao, F ;
Hayashi, F .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1997, 123 (04) :441-447
[4]   Decentration and tilt of polymethyl methacrylate, silicone, and acrylic soft intraocular lenses [J].
Hayashi, K ;
Harada, M ;
Hayashi, H ;
Nakao, F ;
Hayashi, F .
OPHTHALMOLOGY, 1997, 104 (05) :793-798
[5]   THE CORRELATION BETWEEN INCISION SIZE AND CORNEAL SHAPE CHANGES IN SUTURELESS CATARACT-SURGERY [J].
HAYASHI, K ;
HAYASHI, H ;
NAKAO, F ;
HAYASHI, F .
OPHTHALMOLOGY, 1995, 102 (04) :550-556
[6]   Anterior capsule contraction and intraocular lens decentration and tilt after hydrogel lens implantation [J].
Hayashi, K ;
Hayashi, H ;
Nakao, F ;
Hayashi, F .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (11) :1294-1297
[7]  
Marcantonio JM, 2000, INVEST OPHTH VIS SCI, V41, P1130
[8]   Adhesiveness of AcrySof to a collagen film [J].
Nagata, T ;
Minakata, A ;
Watanabe, I .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (03) :367-370
[9]   Speed of capsular bend formation at the optic edge of acrylic, silicone, and poly(methyl methacrylate) lenses [J].
Nishi, O ;
Nishi, K ;
Akura, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (03) :431-437
[10]   Adhesion of lens capsule to intraocular lenses of polymethylmethacrylate, silicone, and acrylic foldable materials: an experimental study [J].
Oshika, T ;
Nagata, T ;
Ishii, Y .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (05) :549-553