Assessment of large single-fraction, low-energy X-ray dose with radiochromic film

被引:14
作者
Butson, MJ
Cheung, T
Yu, PKN
Metcalfe, PE
机构
[1] Illawarra Canc Care Ctr, Dept Med Phys, Wollongong, NSW 2500, Australia
[2] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon Tong, Hong Kong, Peoples R China
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2000年 / 46卷 / 04期
关键词
radiochromic film; in vivo dosimetry; single-fraction irradiation; X-rays; low energy;
D O I
10.1016/S0360-3016(99)00465-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the accuracy of in vivo dosimetry using radiochromic film for large single-fraction, low-energy irradiations. Methods and Materials: Gafchromic MD-55-2 radiochromic film and LiF thermoluminescent dosimeters (TLDs) were placed in vivo on 25 patients to ascertain their effectiveness for assessment of dose. All patients received 10 Gy single fractions at energies ranging from 100 kVp (half-value layer [HVL] = 3.5 mm Al) up to 250 kVp (HVL = 2.3 mm Cu), Effects of small air gaps were also investigated using LiF TLDs and radiochromic film. Results: Radiochromic him adequately measured applied dose for 25 patients in vivo with a standard deviation of 5.5% from prescribed dose. LiF TLDs recorded a standard deviation of 4.1% from measured to applied dose, Small air gaps which can be created under the film or TLDs during in vivo dosimetry were shown to have a measurable but minimal effect on results for gaps less than 5 mm, Conclusions: Gafchromic film has adequately measured applied dose in vivo at low energy for large 10 Gy single-fraction irradiation. (C) 2000 Elsevier Science Inc.
引用
收藏
页码:1071 / 1075
页数:5
相关论文
共 13 条
[1]   Measurement of off-axis and peripheral skin dose using radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) :2647-2650
[2]  
Carolan M. G., 1997, Australasian Physical and Engineering Sciences in Medicine, V20, P183
[3]  
Ferguson S, 1995, Med Dosim, V20, P161, DOI 10.1016/0958-3947(95)00009-L
[4]   GafChromic MD-55: Investigated as a precision dosimeter [J].
Klassen, NV ;
van der Zwan, L ;
Cygler, J .
MEDICAL PHYSICS, 1997, 24 (12) :1924-1934
[5]   TLD extrapolation for skin dose determination in vivo [J].
Kron, T ;
Butson, M ;
Hunt, F ;
Denham, J .
RADIOTHERAPY AND ONCOLOGY, 1996, 41 (02) :119-123
[6]   X-RAY SURFACE DOSE MEASUREMENTS USING TLD EXTRAPOLATION [J].
KRON, T ;
ELLIOT, A ;
WONG, T ;
SHOWELL, G ;
CLUBB, B ;
METCALFE, P .
MEDICAL PHYSICS, 1993, 20 (03) :703-711
[7]   Novel radiochromic films for clinical dosimetry [J].
McLaughlin, WL ;
Puhl, JM ;
AlSheikhly, M ;
Christou, CA ;
Miller, A ;
Kovacs, A ;
Wojnarovits, L ;
Lewis, DF .
RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) :263-268
[8]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[9]   Dosimetric characteristics of an improved radiochromic film [J].
Meigooni, AS ;
Sanders, MF ;
Ibbott, GS ;
Szeglin, SR .
MEDICAL PHYSICS, 1996, 23 (11) :1883-1888
[10]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115