Quality control aspects of the Philips multileaf collimator

被引:18
作者
Hounsell, AR
Jordan, TJ
机构
[1] North Western Medical Physics, Christie Hospital NHS Trust, M20 4BX, Manchester
关键词
multileaf collimators; quality control;
D O I
10.1016/S0167-8140(97)00100-X
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Linear accelerators equipped with multileaf collimators (MLCs) are becoming more common and are widely available from most commercial manufacturers. There is a need to ensure they retain their commissioning specification using a preventative maintenance and quality control (QC) programme. This paper considers the design criteria of the Philips MLC which are important to the production of a comprehensive quality control programme. Materials and methods: The specific QC problems related to MLCs are identified as the positional accuracy of the leaves and their relationship to the back-up collimators, leakage considerations, the relationship of X-ray to light field and the influence of gravity on the positioning and leakage characteristics of the leaves. These problems are considered in relation to the general design considerations of the MLC, and methods of performing routine quality control checks are discussed. Results and conclusions: The introduction of MLCs into clinical use results in new QC procedures being developed but it can be concluded that for the Philips MLC only an extra 30 min of QC time is needed per month and that its use has added little to the general down-time of this department. (C) 1997 Elsevier Science Ireland Ltd.
引用
收藏
页码:225 / 233
页数:9
相关论文
共 14 条
[1]  
*AAPM 40, 1994, MED PHYS, V21, P581
[2]  
*AAPM 4K, 1994, MED PHYS, V21, P1093
[3]   CLINICAL DOSIMETRY FOR IMPLEMENTATION OF A MULTILEAF COLLIMATOR [J].
BOYER, AL ;
OCHRAN, TG ;
NYERICK, CE ;
WALDRON, TJ ;
HUNTZINGER, CJ .
MEDICAL PHYSICS, 1992, 19 (05) :1255-1261
[4]   CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM [J].
GALVIN, JM ;
SMITH, AR ;
LALLY, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :181-192
[5]   COMPUTER-ASSISTED GENERATION OF MULTI-LEAF COLLIMATOR SETTINGS FOR CONFORMATION THERAPY [J].
HOUNSELL, AR ;
SHARROCK, PJ ;
MOORE, CJ ;
SHAW, AJ ;
WILKINSON, JM ;
WILLIAMS, PC .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (772) :321-326
[6]   DOSIMETRIC CHARACTERISTICS OF A COMMERCIAL MULTILEAF COLLIMATOR [J].
HUQ, MS ;
YU, Y ;
CHEN, ZP ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1995, 22 (02) :241-247
[7]  
*I PHYS SCI MED, 1988, 54 IPSM
[8]  
International Electrotechnical Commission, 1989, 977 IEC
[9]  
JOHANSSON KA, 1993, RADIOTHERAPY PHYSICS, P77
[10]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251