Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: An in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation

被引:76
作者
Petersen, HH [1 ]
Chen, X [1 ]
Pietersen, A [1 ]
Svendsen, JH [1 ]
Haunso, S [1 ]
机构
[1] Natl Univ Hosp, Med Dept B, DK-2100 Copenhagen O, Denmark
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2000年 / 23卷 / 01期
关键词
radiofrequency energy; catheter ablation; cooled tip electrode; ventricular arrhythmias; irrigated electrode;
D O I
10.1111/j.1540-8159.2000.tb00644.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The limited success rate of radiofrequency catheter ablation in patients with ventricular tachycardias related to structural heart disease may be increased by enlarging the lesion size. Irrigated tip catheter ablation is a new method for enlarging the size of the lesion. If was introduced in the power-controlled mode with high power and high infusion rate, and is associated with an increased risk of crater formation, which is related to high tissue temperatures. The present study explored the tissue temperatures during temperature-controlled irrigated tip ablation, comparing it with standard temperature-con trolled ablation and power-controlled irrigated tip ablation. In vitro strips of porcine left ventricular myocardium were ablated. Temperature-controlled irrigated tip ablation at target temperatures 60 degrees C, 70 degrees C, and 80 degrees C with infusion of 1 mt saline/min were compared with standard temperature-controlled ablation at 70 degrees C and power-controlled irrigated tip ablation at 40 W, and infusion of 20 mL/min. Lesion size and tissue temperatures were significantly higher during all modes of irrigated tip ablation compared with standard temperature-controlled ablation (P < 0.05). Lesion volume correlated positively with tissue temperature (r = 0.87). The maximum recorded tissue temperature was always 1 mm from the ablation electrode and was 67 +/- 4 degrees C for standard ablation and 93 +/- 6 degrees C, 99 +/- 6 degrees C, and 115 +/- 13 degrees C for temperature-controlled irrigated tip ablation at 60 degrees C, 70 degrees C, and 80 degrees C, respectively and 122 +/- 12 degrees C for power-controlled irrigated tip ablation, which for irrigated tip ablation was significantly higher than tip temperature (P < 0.0001). Crater formation only occurred at tissue temperatures > 100 degrees C. We conclude that irrigated tip catheter ablation increases lesion size and tissue temperatures compared with standard ablation in the temperature-controlled mode at the same or higher target temperatures and in the power-controlled mode. Furthermore, tissue temperature and delivered power are the best indicators of lesion volume during temperature-controlled ablation. (PACE 2000; 23:8-17).
引用
收藏
页码:8 / 17
页数:10
相关论文
共 19 条
[11]   A COMPUTER-SIMULATION OF RADIOFREQUENCY ABLATION OF THE ENDOCARDIUM [J].
LABONTE, S .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (09) :883-890
[12]   USE OF THE SALINE INFUSION ELECTRODE CATHETER FOR IMPROVED ENERGY DELIVERY AND INCREASED LESION SIZE IN RADIOFREQUENCY CATHETER ABLATION [J].
MITTLEMAN, RS ;
HUANG, SKS ;
DEGUZMAN, WT ;
CUENOUD, H ;
WAGSHAL, AB ;
PIRES, LA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :1022-1027
[13]   COMPARISON OF IN-VIVO TISSUE TEMPERATURE PROFILE AND LESION GEOMETRY FOR RADIOFREQUENCY ABLATION WITH A SALINE-IRRIGATED ELECTRODE VERSUS TEMPERATURE CONTROL IN A CANINE THIGH MUSCLE PREPARATION [J].
NAKAGAWA, H ;
YAMANASHI, WS ;
PITHA, JV ;
ARRUDA, M ;
WANG, XZ ;
OHTOMO, K ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1995, 91 (08) :2264-2273
[14]   Why a large tip electrode makes a deeper radiofrequency lesion: Effects of increase in electrode cooling and electrode-tissue interface area [J].
Otomo, K ;
Yamanashi, WS ;
Tondo, C ;
Antz, M ;
Bussey, J ;
Pitha, JV ;
Arruda, M ;
Nakagawa, H ;
Wittkampf, FHM ;
Lazzara, R ;
Jackman, WM .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (01) :47-54
[15]  
PACKER DL, 1997, PACE, V20, P1565
[16]   Lesion dimensions during temperature-controlled radiofrequency catheter ablation of left ventricular porcine myocardium - Impact of ablation site, electrode size, and convective cooling [J].
Petersen, HH ;
Chen, X ;
Pietersen, A ;
Svendsen, JH ;
Haunso, S .
CIRCULATION, 1999, 99 (02) :319-325
[17]   Temperature-controlled irrigated tip radiofrequency catheter ablation: Comparison of in vivo and in vitro lesion dimensions for standard catheter and irrigated tip catheter with minimal infusion rate [J].
Petersen, HH ;
Chen, X ;
Pietersen, A ;
Svendsen, JH ;
Haunso, S .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (04) :409-414
[18]  
SKRUMEDA L, 1997, PACE 2, V20, P1206
[19]  
WITTKAMP FH, 1997, PACE 2, V20, P1206