Four-day incubation period for blood culture bottles processed with the Difco ESP blood culture system

被引:30
作者
Doern, GV
Brueggemann, AB
Dunne, WM
Jenkins, SG
Halstead, DC
McLaughlin, JC
机构
[1] HENRY FORD HOSP,DETROIT,MI 48202
[2] UNIV FLORIDA,HLTH SCI CTR,JACKSONVILLE,FL 32209
[3] UNIV NEW MEXICO,HLTH SCI CTR,ALBUQUERQUE,NM 87106
[4] BAPTIST MED CTR WOLFSON CHILDRENS HOSP,JACKSONVILLE,FL 32207
关键词
D O I
10.1128/JCM.35.5.1290-1292.1997
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Blood culture records from 1991 to 1995 from five U.S. medical centers all using the Difco ESP continuous monitoring blood culture system were reviewed retrospectively. Among a total of 7,362 isolates of bacteria and yeasts, only 0.1% of possibly significant isolates would have been missed had blood cultures been routinely incubated for 4 days instead of the 5 days recommended by the manufacturer, Conversely, numerous contaminants, detected only on day 5, would have been eliminated by a 4-day incubation period.
引用
收藏
页码:1290 / 1292
页数:3
相关论文
共 11 条
[1]   TIME TO DETECTION OF POSITIVE BACT ALERT BLOOD CULTURES AND LACK OF NEED FOR ROUTINE SUBCULTURE OF 5-DAY TO 7-DAY NEGATIVE CULTURES [J].
HARDY, DJ ;
HULBERT, BB ;
MIGNEAULT, PC .
JOURNAL OF CLINICAL MICROBIOLOGY, 1992, 30 (10) :2743-2745
[2]   SUBCULTURES OF BACTEC-POSITIVE BUT GRAM OR ACRIDINE-ORANGE STAIN-NEGATIVE NR 6A AND 7A BLOOD CULTURE BOTTLES ARE UNNECESSARY [J].
HARRELL, LJ ;
MIRRETT, S ;
RELLER, LB .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1994, 20 (03) :121-125
[3]   DETECTION OF BACTEREMIA BY DIFCO ESP BLOOD CULTURE SYSTEM [J].
MORELLO, JA ;
LEITCH, C ;
NITZ, S ;
DYKE, JW ;
ANDRUSZEWSKI, M ;
MAIER, G ;
LANDAU, W ;
BEARD, MA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (03) :811-818
[4]   MULTICENTER CLINICAL-EVALUATION OF A CONTINUOUS MONITORING BLOOD CULTURE SYSTEM USING FLUORESCENT-SENSOR TECHNOLOGY (BACTEC-9240) [J].
NOLTE, FS ;
WILLIAMS, JM ;
JERRIS, RC ;
MORELLO, JA ;
LEITCH, CD ;
MATUSHEK, S ;
SCHWABE, LD ;
DORIGAN, F ;
KOCKA, FE .
JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (03) :552-557
[5]   CONTROLLED CLINICAL-EVALUATION OF BACTEC PLUS AEROBIC/F AND BACT/ALERT AEROBIC FAN BOTTLES FOR DETECTION OF BLOOD-STREAM INFECTIONS [J].
POHLMAN, JK ;
KIRKLEY, BA ;
EASLEY, KA ;
BASILLE, BA ;
WASHINGTON, JA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (11) :2856-2858
[6]   EVALUATION OF BACTEC-9240 BLOOD CULTURE SYSTEM BY USING HIGH-VOLUME AEROBIC RESIN MEDIA [J].
SCHWABE, LD ;
THOMSON, RB ;
FLINT, KK ;
KOONTZ, FP .
JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (09) :2451-2453
[7]   COMPARISON OF BACTEC-9240 AND BACT/ALERT BLOOD CULTURE SYSTEMS IN AN ADULT HOSPITAL [J].
SMITH, JA ;
BRYCE, EA ;
NGUIYEN, JH ;
ROBERTS, FJ .
JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (07) :1905-1908
[8]   BACT/ALERT - AN AUTOMATED COLORIMETRIC MICROBIAL DETECTION SYSTEM [J].
THORPE, TC ;
WILSON, ML ;
TURNER, JE ;
DIGUISEPPI, JL ;
WILLERT, M ;
MIRRETT, S ;
RELLER, LB .
JOURNAL OF CLINICAL MICROBIOLOGY, 1990, 28 (07) :1608-1612
[9]   RECOVERY OF CLINICALLY IMPORTANT MICROORGANISMS FROM THE BACT/ALERT BLOOD CULTURE SYSTEM DOES NOT REQUIRE TESTING FOR 7 DAYS [J].
WILSON, ML ;
MIRRETT, S ;
RELLER, LB ;
WEINSTEIN, MP ;
REIMER, LG .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1993, 16 (01) :31-34
[10]   CONTROLLED EVALUATION OF BACT ALERT STANDARD ANAEROBIC AND FAN ANAEROBIC BLOOD CULTURE BOTTLES FOR THE DETECTION OF BACTEREMIA AND FUNGEMIA [J].
WILSON, ML ;
WEINSTEIN, MP ;
MIRRETT, S ;
REIMER, LG ;
FELDMAN, RJ ;
CHUARD, CR ;
RELLER, LB .
JOURNAL OF CLINICAL MICROBIOLOGY, 1995, 33 (09) :2265-2270