Effects of a heat and moisture exchanger and a heated humidifier on respiratory mucus in patients undergoing mechanical ventilation

被引:46
作者
Nakagawa, NK [1 ]
Macchione, M
Petrolino, HMS
Guimaraes, ET
King, M
Saldiva, PHN
Lorenzi, G
机构
[1] Univ Sao Paulo, Fac Med, BR-05508 Sao Paulo, Brazil
[2] Hosp Sirio Libanes, Sao Paulo, Brazil
[3] Univ Fed Sao Paulo, Sao Paulo, Brazil
[4] Pulm Res Grp, Edmonton, AB, Canada
关键词
mechanical ventilation; humidification; heat and moisture exchanger; mucus; rheology; mucociliary clearance; cough;
D O I
10.1097/00003246-200002000-00004
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the effects of a heat and moisture exchanger and a heated humidifier on respiratory mucus and transportability by cilia and cough in patients undergoing invasive mechanical ventilation (up to 72 hrs). Design: Prospective, randomized, clinical study. Setting: General intensive care unit and university research laboratory. Patients: A total of 32 consecutive patients with acute respiratory failure, who were intubated and mechanically ventilated in the intensive care unit setting, were enrolled in the study. interventions: Patients were randomly assigned to receive as a humidifying system a heat and moisture exchanger (HME) or heated humidified water (HHW) at the onset of mechanical ventilation (time 0). Respiratory mucus samples were collected by suction using a sterile technique at time 0, 24, 48, and 72 hrs of mechanical ventilation. Measurements and Main Results: Eleven patients were excluded from this study because of either extubation or death before 72 hrs of mechanical ventilation, leaving 12 patients in the HME group and nine patients in the HHW group. Ventilatory variables including minute volume, mean airway pressure, positive end-expiratory pressure, Fio(2), as well as Pao(2)/Fio(2) ratio, fluid balance (last 6 hrs), furosemide, and inotrope administration (last 4 hrs) were recorded. In vitro mucus transportability by cilia was evaluated on the mucus-depleted frog palate model, and the results were expressed as the mucus transport rate. Cough clearance (an estimation of the interaction between the flow of air and the mucus lining the bronchial walls) was measured using a simulated cough machine, the results being expressed in millimeters. Mucus wettability was measured by the contact angle between a mucus sample drop and a flat glass surface. Mucus theologic properties (mechanical impedance [log G*] and the ratio between viscosity and elasticity [tan delta]) were measured using a magnetic microrheometer at 1 and 100 cGy/sec deformation frequency. The two humidification groups were comparable in terms of the Acute Physiology and Chronic Health Evaluation II score, age, gender, ventilatory variables, fluid balance, use of inotropes, and furosemide. Conclusion: Ours results indicate that air humidification with either HME or HHW at 32 degrees C (89.6 degrees F) has similar effects on mucus theologic properties, contact angle, and transportability by cilia in patients undergoing mechanical ventilation, except for transportability by cough, which diminished after 72 hrs of mechanical ventilation in the HME group (p = .0441).
引用
收藏
页码:312 / 317
页数:6
相关论文
共 36 条
[1]  
ACKERMAN MH, 1985, HEART LUNG, V14, P505
[2]   EFFECTS OF DRY ANESTHETIC GASES ON TRACHEOBRONCHIAL CILIATED EPITHELIUM [J].
CHALON, J ;
MALEBRANCHE, J ;
LOEW, DAY .
ANESTHESIOLOGY, 1972, 37 (03) :338-+
[3]   USE OF HEAT AND MOISTURE EXCHANGING (HME) FILTERS IN MECHANICALLY VENTILATED ICU PATIENTS - INFLUENCE ON AIRWAY FLOW-RESISTANCE [J].
CHIARANDA, M ;
VERONA, L ;
PINAMONTI, O ;
DOMINIONI, L ;
MINOJA, G ;
CONTI, G .
INTENSIVE CARE MEDICINE, 1993, 19 (08) :462-466
[4]   PERFORMANCE OF A HYDROPHOBIC HEAT AND MOISTURE EXCHANGER AT DIFFERENT AMBIENT-TEMPERATURES [J].
CROCI, M ;
ELENA, A ;
SOLCA, M .
INTENSIVE CARE MEDICINE, 1993, 19 (06) :351-352
[5]  
DREYFUSS D, 1995, AM J RESP CRIT CARE, V151, P986
[6]   TEMPERATURE, HUMIDITY AND MUCUS FLOW IN INTUBATED TRACHEA [J].
FORBES, AR .
BRITISH JOURNAL OF ANAESTHESIA, 1974, 46 (01) :29-34
[7]   CONTAMINATION CONTROL IN LONG-TERM VENTILATION - A CLINICAL-STUDY USING A HEAT-EXCHANGING AND MOISTURE-EXCHANGING FILTER [J].
GALLAGHER, J ;
STRANGEWAYS, JEM ;
ALLTGRAHAM, J .
ANAESTHESIA, 1987, 42 (05) :476-481
[8]   HEAT AND MOISTURE EXCHANGERS AND BREATHING FILTERS [J].
HEDLEY, RM ;
ALLTGRAHAM, J .
BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (02) :227-236
[9]   GASTRIC-MUCOSAL BARRIER - STABILIZATION OF HYDROPHOBIC LINING TO THE STOMACH BY MUCUS [J].
HILLS, BA .
AMERICAN JOURNAL OF PHYSIOLOGY, 1985, 249 (03) :G342-G349
[10]   AN EVALUATION OF THE MICROBIAL RETENTION PERFORMANCE OF 3 VENTILATOR-CIRCUIT FILTERS [J].
HOLTON, J ;
WEBB, AR .
INTENSIVE CARE MEDICINE, 1994, 20 (03) :233-237