A computational study of the effects of inlet guide vanes on the performance of a centrifugal blood pump

被引:23
作者
Chan, WK [1 ]
Wong, YW [1 ]
Yu, SCM [1 ]
Chua, LP [1 ]
机构
[1] Nanyang Technol Univ, Sch MPE, Singapore 639798, Singapore
关键词
centrifugal blood pump; inlet guide vanes; numerical simulation;
D O I
10.1046/j.1525-1594.2002.06917.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This article presents computational studies on the effects of inlet guide vanes (IGVs) on the flow pattern and shear stress in a centrifugal blood pump. The effect of IGVs is to introduce a pre-swirl to fluid particles entering the impeller with the intention that the fluid particles will travel along the blade profile. Currently, most commercial centrifugal blood pumps employ straight radial impeller blades that are not hydrodynamically ideal for a good flow pattern within the blade passage. Flow separation and formation of vortices within the blade passage are believed to increase the degree of hemolysis and thrombosis. These are causes for blood clotting that will lead to malfunctioning of ventricular assist devices. Four IGVs of different geometrical profiles have been numerically investigated using a commercial software program CFX-Tascflow. The pump is operated at 2,000 rpm, and the results revealed that the relative flow patterns in the blade passage have been dramatically altered. The size of the vortices was reduced, and the pressure contours indicated a gradual rise from the impeller leading edge to the trailing edge. However, inclusion of IGV causes a drop in the pressure head generated. Higher frictional losses are incurred as fluid particles passed through the IGV. In addition, the IGV modifies the inlet velocity triangles, and this also contributes to a drop in the pressure head generated that is consistent with Euler's pump theory. The change in the flow patterns and the gradual variation of the pressure contours have led to lower shear stress within the blade passages as compared to the case without IGVs.
引用
收藏
页码:534 / 542
页数:9
相关论文
共 12 条
[1]   RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
TSUKIYA, T ;
NISHIMURA, K ;
PARK, CH ;
NAKAZEKI, T .
ARTIFICIAL ORGANS, 1995, 19 (07) :631-634
[2]  
AKAMATSU T, 2000, ROTARY BLOOD PUMPS N, P35
[3]   Prototype continuous flow ventricular assist device supported on magnetic bearings [J].
Allaire, PE ;
Kim, HC ;
Maslen, EH ;
Olsen, DB ;
Bearnson, GB .
ARTIFICIAL ORGANS, 1996, 20 (06) :582-590
[4]   A magnetically suspended and hydrostatically stabilized centrifugal blood pump [J].
Hart, RM ;
Filipenco, VG ;
Kung, RTV .
ARTIFICIAL ORGANS, 1996, 20 (06) :591-596
[5]  
Ikeda T, 1996, ARTIF ORGANS, V20, P132
[6]  
MULDER MM, 1993, ASME ADV BIOENGINEER, V26, P443
[7]   Experimental and computational studies of the relative flow field in a centrifugal blood pump [J].
Ng, BTH ;
Chan, WK ;
Yu, SCM ;
Li, HD .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 2000, 28 (1-2) :119-125
[8]  
NOON GP, 1990, T AM SOC ART INT ORG, V36, P616
[9]  
Qian K X, 1990, ASAIO Trans, V36, P33
[10]   LOW HEMOLYSIS PULSATILE IMPELLER PUMP - DESIGN CONCEPTS AND EXPERIMENTAL RESULTS [J].
QIAN, KX .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (06) :478-481