Experimental and computational studies of the relative flow field in a centrifugal blood pump

被引:8
作者
Ng, BTH [1 ]
Chan, WK [1 ]
Yu, SCM [1 ]
Li, HD [1 ]
机构
[1] Nanyang Technol Univ, Sch Mech & Prod Engn, Thermal & Fluids Engn Div, Singapore 639798, Singapore
关键词
impeller design; rotary blood pump; computational fluid dynamics and flow visualization;
D O I
10.1615/CritRevBiomedEng.v28.i12.200
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The relative flow field within the impeller passage of a centrifugal blood pump had been examined using flow visualization technique and computational fluid dynamics. It was found that for a seven-blade radial impeller design, the required now rate and static pressure rise across the pump could be achieved but the flow field within the blades was highly undesirable. Two vortices were observed near the: suction side and these could lead to thrombus formation. preliminary results presented in this article are part of our overall effort to minimize undesirable flow patterns such flow separation and high shear stress regions within the centrifugal blood pump. This will facilitate the future progress in developing a long-term clinically effective blood pump.
引用
收藏
页码:119 / 125
页数:7
相关论文
共 4 条
[1]   RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
TSUKIYA, T ;
NISHIMURA, K ;
PARK, CH ;
NAKAZEKI, T .
ARTIFICIAL ORGANS, 1995, 19 (07) :631-634
[2]   Prototype continuous flow ventricular assist device supported on magnetic bearings [J].
Allaire, PE ;
Kim, HC ;
Maslen, EH ;
Olsen, DB ;
Bearnson, GB .
ARTIFICIAL ORGANS, 1996, 20 (06) :582-590
[3]   SEAL-LESS CENTRIFUGAL BLOOD PUMP WITH MAGNETICALLY SUSPENDED ROTOR - ROT-A-FLOT [J].
MENDLER, N ;
PODECHTL, F ;
FEIL, G ;
HILTMANN, P ;
SEBENING, F .
ARTIFICIAL ORGANS, 1995, 19 (07) :620-624
[4]   Quantitative visualization of flow through a centrifugal blood pump: Effect of washout holes [J].
Nishida, M ;
Yamane, T ;
Orita, T ;
Asztalos, B ;
Clarke, H .
ARTIFICIAL ORGANS, 1997, 21 (07) :720-729