This study was designed to determine the changes in precision of the ultrasound parameters speed of sound (SOS), broadband ultrasound attenuation (BUA), and stiffness index (SI) as a function of bone quality. The instrument used in this investigation was the LUNAR Achilles. Of the 608 female patients who had paired measurement with repositioning, 200 had t scores greater than or equal to-1 and 408 had t scores less than or equal to-2.5, thus rendering a normal and a grossly abnormal group for comparison of precision results. It was found that the median precision error (CV%) for BUA was 1.99% for normal bone and 1.44% for abnormal bone (P = 0.02). No significant difference was obtained between median precision errors in normal and abnormal bone for the parameters SOS and SI, which were 0.23% and 0.19%, and 2.15% and 2.02%, respectively. In the interquartile range, the mean precision errors for SOS and BUA were significantly different in normal and abnormal bone: SOS was 0.25% and 0.21%, respectively, and BUA was 2.31% and 1.85%, respectively. No significant change was registered for SI. Precision error appears to decrease slightly at lower values for calcaneal variables.