Effect of sensing system on the incidence of myopotential oversensing during bradycardia pacing in implantable cardioverter-defibrillators

被引:12
作者
Mann, DE [1 ]
Otto, L [1 ]
Kelly, PA [1 ]
Reiter, MJ [1 ]
机构
[1] Univ Colorado, Hlth Sci Ctr, Div Cardiol, Dept Med, Denver, CO 80262 USA
关键词
D O I
10.1016/S0002-9149(00)00776-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We compared the incidence of oversensing of myopotentials during bradycardia pacing in implantable cardioverter-defibrillators from 3 different manufacturers, with 3 different types of sensing systems. The Ventritex sensing system, which uses maximum gain continuously during bradycardia pacing, had a higher incidence of oversensing (24% of patients) than the sensing systems of CPI (3%) or Medtronic (0%), suggesting that the Ventritex sensing system may be unsuitable for use in a dual-chamber defibrillator.
引用
收藏
页码:1380 / +
页数:4
相关论文
共 7 条
[1]   OVERSENSING DURING VENTRICULAR PACING IN PATIENTS WITH A 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
KELLY, PA ;
MANN, DE ;
DAMLE, RS ;
REITER, MJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (07) :1531-1534
[2]   Comparison of oversensing during bradycardia pacing in two types of implantable cardioverter-defibrillator systems [J].
Mann, DE ;
Damle, RS ;
Kelly, PA ;
Landers, M ;
Otto, L ;
Reiter, MJ .
AMERICAN HEART JOURNAL, 1998, 136 (04) :658-663
[3]  
Newcombe RG, 1998, STAT MED, V17, P857, DOI 10.1002/(SICI)1097-0258(19980430)17:8<857::AID-SIM777>3.0.CO
[4]  
2-E
[5]   Inappropriate discharge of an implantable cardioverter defibrillator during atrial flutter and intermittent ventricular antibradycardia pacing [J].
Peters, W ;
Kowallik, P ;
Wittenberg, G ;
Scholl, C ;
Meesmann, M .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (10) :1167-1174
[6]   Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators [J].
Reiter, MJ ;
Mann, DE .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) :542-558
[7]   Noise detection during bradycardia pacing with a hybrid nonthoracotomy implantable cardioverter defibrillator system: Incidence and clinical significance [J].
Rosenthal, ME ;
Paskman, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (07) :1380-1386