Magnetic resonance renography: Optimisation of pulse sequence parameters and Gd-DTPA dose, and comparison with radionuclide renography

被引:57
作者
Taylor, J [1 ]
Summers, PE [1 ]
Keevil, SF [1 ]
Saks, AM [1 ]
Diskin, J [1 ]
Hilton, PJ [1 ]
Ayers, AB [1 ]
机构
[1] UNITED MED & DENT SCH GUYS & ST THOMASS HOSP,GUYS & ST THOMAS MAGNET RESONANCE CTR,LONDON SE1 7EH,ENGLAND
关键词
magnetic resonance imaging; kidneys; gadolinium; enhancement;
D O I
10.1016/S0730-725X(97)00034-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to assess the feasibility of magnetic resonance renography (MRR) using gadolinium diethylenetriaminepentaacetic acid (Gd-DTPA) in comparison with conventional radionuclide renography (RR) using technetium-99m-DTPA (Tc-99m-DTPA). MRR has many advantages over RR, including lack of ionising radiation, increased spatial resolution, and visible background anatomy. By optimising the pulse sequence, we developed an MRR protocol in which signal intensity is linear with Gd-DTPA concentration over a clinically relevant range. Twenty-nine patients and a volunteer were studied using this protocol. Magnetic resonance renography was performed using three different doses of Gd-DTPA: 0.1 mmol kg(-1) (n = 13), 0.05 mmol kg(-1) (n = 7), and 0.025 mmol kg(-1) (n = 9). Each patient was also assessed using radionuclide renography. The resulting renograms were assessed in terms of time to peak signal intensity, signal decrease after peak, and kidney function ratios calculated from both the areas underneath and the slopes of the uptake curves. We have shown that the MR renograms obtained using-low dose Gd-DTPA correlate best with the radionuclide renograms. Remaining discrepancies may be explained by variations in the injection procedures (hence in arterial input functions) and the limited coverage of the three MRR slices compared to the whole body projection of RR. Furthermore, at high local concentrations, signal becomes independent of T-1 and is dominated by T-2(*). (C) 1997 Elsevier Science Inc.
引用
收藏
页码:637 / 649
页数:13
相关论文
共 26 条
[1]  
BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
[2]  
BROOKES JA, 1995, 3 ANN M SOC MAGN RES, P520
[3]   ACUTE TUBULAR-NECROSIS - USE OF GADOLINIUM-DTPA AND FAST MR IMAGING TO EVALUATE RENAL-FUNCTION IN THE RABBIT [J].
CARVLIN, MJ ;
ARGER, PH ;
KUNDEL, HL ;
AXEL, L ;
DOUGHERTY, L ;
KASSAB, EA ;
MOORE, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (03) :488-495
[4]   DYNAMIC GD-DTPA - ENHANCED MR IMAGING OF THE KIDNEY - EXPERIMENTAL RESULTS [J].
CHOYKE, PL ;
FRANK, JA ;
GIRTON, ME ;
INSCOE, SW ;
CARVLIN, MJ ;
BLACK, JL ;
AUSTIN, HA ;
DWYER, AJ .
RADIOLOGY, 1989, 170 (03) :713-720
[5]   FUNCTIONAL MR OF THE KIDNEY [J].
FRANK, JA ;
CHOYKE, PL ;
AUSTIN, HA ;
GIRTON, ME .
MAGNETIC RESONANCE IN MEDICINE, 1991, 22 (02) :319-323
[6]  
FRANK JA, 1992, 11 ANN M SOC MAGN RE, P121
[7]   FAST MR-IMAGING - TECHNIQUES AND CLINICAL-APPLICATIONS [J].
HAACKE, EM ;
TKACH, JA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 155 (05) :951-964
[8]   SLICE SELECTION AND T1 CONTRAST IN FLASH NMR IMAGING [J].
HANICKE, W ;
MERBOLDT, KD ;
FRAHM, J .
JOURNAL OF MAGNETIC RESONANCE, 1988, 77 (01) :64-74
[9]  
HATHOUT G, 1996, 4 ANN M SOC MAGN RES, P1574
[10]  
HAUSER W, 1970, RADIOLOGY, V94, P697