Feasibility and limitations of bulk density assignment in MRI for head and neck IMRT treatment planning

被引:57
作者
Chin, Alexander L. [1 ]
Lin, Alexander [2 ]
Anamalayil, Shibu [2 ]
Teo, Boon-Keng Kevin [2 ]
机构
[1] Univ Penn, Perelman Sch Med, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2014年 / 15卷 / 05期
关键词
MRI; dose calculation; IMRT; treatment planning; INTENSITY-MODULATED RADIOTHERAPY; ULTRASHORT ECHO TIME; NASOPHARYNGEAL CARCINOMA; MAGNETIC-SUSCEPTIBILITY; GEOMETRIC DISTORTION; PROSTATE-CANCER; CT IMAGES; T MRI; IRRADIATION; SEQUENCES;
D O I
10.1120/jacmp.v15i5.4851
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Head and neck cancers centered at the base of skull are better visualized on MRI than on CT. The purpose of this investigation was to investigate the accuracy of bulk density assignment in head and neck intensity-modulated radiation therapy (IMRT) treatment plan optimization. Our study investigates dose calculation differences between density-assigned MRI and CT, and identifies potential limitations related to dental implants and MRI geometrical distortion in the framework of MRI-only-based treatment planning. Bulk density assignment was performed and applied onto MRI to generate three MRI image sets with increasing levels of heterogeneity for seven patients: 1) MRIW: all water-equivalent; 2) MRIW+B: included bone with density of 1.53 g/cm(3); and 3) MRIW+B+A: included bone and air. Using identical planning and optimization parameters, MRI-based IMRT plans were generated and compared to corresponding, forward-calculated, CT-based plans on the basis of target coverage, isodose distributions, and dose-volume histograms (DVHs). Phantom studies were performed to assess the magnitude and spatial dependence of MRI geometrical distortion. MRIW-based dose calculations overestimated target coverage by 16.1%. Segmentation of bone reduced differences to within 2% of the coverage area on the CT-based plan. Further segmentation of air improved conformity near air-tissue interfaces. Dental artifacts caused substantial target coverage overestimation even on MRIW+B+A. Geometrical distortion was less than 1 mm in an imaging volume 20 x 20 x 20 cm(3) around scanner isocenter, but up to 4 mm at 17 cm lateral to isocenter. Bulk density assignment in the framework of MRI-only IMRT head and neck treatment planning is a feasible method with certain limitations. Bone and teeth account for the majority of density heterogeneity effects. While soft tissue is well visualized on MRI compared to CT, dental implants may not be visible on MRI and must be identified by other means and assigned appropriate density for accurate dose calculation. Far off-center geometrical distortion of the body contour near the shoulder region is a potential source of dose calculation inaccuracy.
引用
收藏
页码:100 / 111
页数:12
相关论文
共 33 条
[1]   Characterization, prediction, and correction of geometric distortion in 3 T MR images [J].
Baldwin, Lesley N. ;
Wachowicz, Keith ;
Thomas, Steven D. ;
Rivest, Ryan ;
Fallone, B. Gino .
MEDICAL PHYSICS, 2007, 34 (02) :388-399
[2]   MRI-based treatment planning for radiotherapy: Dosimetric verification for prostate IMRT [J].
Chen, LL ;
Price, RA ;
Wang, L ;
Li, JS ;
Qin, LH ;
McNeeley, S ;
Ma, CMC ;
Freedman, GM ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (02) :636-647
[3]   Investigation of MR image distortion for radiotherapy treatment planning of prostate cancer [J].
Chen, Z ;
Ma, CM ;
Paskalev, K ;
Li, J ;
Yang, J ;
Richardson, T ;
Palacio, L ;
Xu, X ;
Chen, L .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06) :1393-1403
[4]   MR-driven metal artifact reduction in PET/CT [J].
Delso, G. ;
Wollenweber, S. ;
Lonn, A. ;
Wiesinger, F. ;
Veit-Haibach, P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (07) :2267-2280
[5]   An Atlas-Based Electron Density Mapping Method for Magnetic Resonance Imaging (MRI)-Alone Treatment Planning and Adaptive MRI-Based Prostate Radiation Therapy [J].
Dowling, Jason A. ;
Lambert, Jonathan ;
Parker, Joel ;
Salvado, Olivier ;
Fripp, Jurgen ;
Capp, Anne ;
Wratten, Chris ;
Denham, James W. ;
Greer, Peter B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (01) :E5-E11
[6]   A simulation of MRI based dose calculations on the basis of radiotherapy planning CT images [J].
Eilertsen, Karsten ;
Vestad, Line Nilsen Tor Arne ;
Geier, Oliver ;
Skretting, Arne .
ACTA ONCOLOGICA, 2008, 47 (07) :1294-1302
[7]   Influence of MRI on target volume delineation and IMRT planning in nasopharyngeal carcinoma [J].
Emami, B ;
Sethi, A ;
Petruzzelli, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (02) :481-488
[8]   First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system [J].
Fallone, B. G. ;
Murray, B. ;
Rathee, S. ;
Stanescu, T. ;
Steciw, S. ;
Vidakovic, S. ;
Blosser, E. ;
Tymofichuk, D. .
MEDICAL PHYSICS, 2009, 36 (06) :2084-2088
[9]   Dental Magnetic Resonance Imaging: Making the Invisible Visible [J].
Idiyatullin, Djaudat ;
Corum, Curt ;
Moeller, Steen ;
Prasad, Had S. ;
Garwood, Michael ;
Nixdorf, Donald R. .
JOURNAL OF ENDODONTICS, 2011, 37 (06) :745-752
[10]   CT substitute derived from MRI sequences with ultrashort echo time [J].
Johansson, Adam ;
Karlsson, Mikael ;
Nyholm, Tufve .
MEDICAL PHYSICS, 2011, 38 (05) :2708-2714