Breath condenser coatings affect measurement of biomarkers in exhaled breath condensate

被引:80
作者
Rosias, P. P.
Robroeks, C. M.
Niemarkt, H. J.
Kester, A. D.
Vernooy, J. H.
Suykerbuyk, J.
Teunissen, J.
Heynens, J.
Hendriks, H. J.
Jobsis, Q.
Dompeling, E.
机构
[1] Maasland Hosp, Dept Paediat, NL-6130 MB Sittard, Netherlands
[2] Univ Hosp Maastricht, Dept Paediat Pulm, Maastricht, Netherlands
[3] Univ Hosp Maastricht, Dept Resp Med, Maastricht, Netherlands
[4] Maastricht Univ, Dept Methodol & Stat, Maastricht, Netherlands
关键词
coating; cytokine; exhaled breath condensate; exhaled markers; isoprostanes; methodology;
D O I
10.1183/09031936.06.00110305
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Exhaled breath condensate collection is not yet standardised and biomarker measurements are often close to lower detection limits. In the current study, it was hypothesised that adhesive properties of different condenser coatings interfere with measurements of eicosanoids and proteins in breath condensate. In vitro, condensate was derived from a collection model using two test solutions (8-isoprostane and albumin) and five condenser coatings (silicone, glass, aluminium, polypropylene and Teflon). In vivo, condensate was collected using these five coatings and the EcoScreen((R)) condenser to measure 8-isoprostane, and three coatings (silicone, glass, EcoScreen((R))) to measure albumin. In vitro, silicone and glass coatings had significantly higher albumin recovery compared with the other coatings. A similar trend was observed for 8-isoprostane recovery. In vivo, median (interquartile range) 8-isoprostane concentrations were significantly higher using silicone (9.2 (18.8) pg.mL(-1)) or glass (3.0 (4.5) pg.mL(-1)) coating, compared with aluminium (0.5 (2.4) pg-mL(-1)), polypropylene (0.5 (0.5) pg-mL(-1)), Teflon (0.5 (0.0) pg-mL(-1)), and EcoScreen((R)) (0.5 (2.0) pg.mL(-1)). Albumin in vivo was mainly detectable using glass coating. In conclusion, a condenser with silicone or glass coating is more efficient for measurement of 8-isoprostane or albumin in exhaled breath condensate, than EcoScreen((R)), aluminium, polypropylene or Teflon. Guidelines for exhaled breath condensate standardisation should include the most valid condenser coating to measure a specific biomarker.
引用
收藏
页码:1036 / 1041
页数:6
相关论文
共 26 条
[11]   Physicochemical characterization of cross-linked human serum albumin dimer and its synthetic heme hybrid as an oxygen carrier [J].
Komatsu, T ;
Oguro, Y ;
Teramura, Y ;
Takeoka, S ;
Okai, J ;
Anraku, M ;
Otagiri, M ;
Tsuchida, E .
BIOCHIMICA ET BIOPHYSICA ACTA-GENERAL SUBJECTS, 2004, 1675 (1-3) :21-31
[12]   Analysis of growth factors and inflammatory cytokines in exhaled breath condensate from asthmatic children [J].
Leung, TF ;
Wong, GWK ;
Ko, FWS ;
Li, CY ;
Yung, E ;
Lam, CWK ;
Fok, TF .
INTERNATIONAL ARCHIVES OF ALLERGY AND IMMUNOLOGY, 2005, 137 (01) :66-72
[13]   The relation between obesity and asthmatic airway inflammation [J].
Leung, TF ;
Li, CY ;
Lam, CWK ;
Au, CSS ;
Yung, E ;
Chan, IHS ;
Wong, GWK ;
Fok, TF .
PEDIATRIC ALLERGY AND IMMUNOLOGY, 2004, 15 (04) :344-350
[14]   Validation of 8-isoprostane and prostaglandin E2 measurements in exhaled breath condensate [J].
Montuschi, P ;
Ragazzoni, E ;
Valente, S ;
Corbo, G ;
Mondino, C ;
Ciappi, G ;
Ciabattoni, G .
INFLAMMATION RESEARCH, 2003, 52 (12) :502-507
[15]   Analysis of exhaled breath condensate for monitoring airway inflammation [J].
Montuschi, P ;
Barnes, PJ .
TRENDS IN PHARMACOLOGICAL SCIENCES, 2002, 23 (05) :232-237
[16]   Collection and analysis of exhaled breath condensate in humans [J].
Mutlu, GM ;
Garey, KW ;
Robbins, RA ;
Danziger, LH ;
Rubinstein, I .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2001, 164 (05) :731-737
[17]   Reproducibility of oxidative stress biomarkers in breath condensate: are they reliable? [J].
Rahman, I .
EUROPEAN RESPIRATORY JOURNAL, 2004, 23 (02) :183-184
[18]   Exhaled breath condensate:: a space odessey, where no one has gone before... [J].
Rosias, P ;
Robroeks, C ;
Hendriks, J ;
Dompeling, E ;
Jöbsis, Q .
EUROPEAN RESPIRATORY JOURNAL, 2004, 24 (01) :189-190
[19]   Childhood asthma:: Exhaled markers of airway inflammation, asthma control score, and lung function tests [J].
Rosias, PPR ;
Dompeling, E ;
Dentener, MA ;
Pennings, HJ ;
Hendriks, HJE ;
Van Iersel, MPA ;
Jöbsis, Q .
PEDIATRIC PULMONOLOGY, 2004, 38 (02) :107-114
[20]   Exhaled breath condensate in children:: Pearls and pitfalls [J].
Rosias, PPR ;
Dompeling, E ;
Hendriks, HJE ;
Heijnens, JWCM ;
Donckerwolcke, RAMG ;
Jöbsis, Q .
PEDIATRIC ALLERGY AND IMMUNOLOGY, 2004, 15 (01) :4-19