Precision of EBRA-digital software for monitoring implant migration after total hip arthroplasty

被引:60
作者
Wilkinson, JM [1 ]
Hamer, AJ
Elson, RA
Stockley, I
Eastell, R
机构
[1] Univ Sheffield, No Gen Hosp, Div Clin Sci NGHT, Med Sect,Bone Metab Grp, Sheffield S5 7AU, S Yorkshire, England
[2] No Gen Hosp, Dept Orthopaed, Hip & Knee Arthroplasty Unit, Sheffield S5 7AU, S Yorkshire, England
关键词
total hip arthroplasty (THA); EBRA; precision; migration; aseptic loosening;
D O I
10.1054/arth.2002.34530
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We assessed the precision of the EBRA-Digital software (EBRA, University of Innsbruck, Innsbruck, Austria) for measuring implant migration after total hip arthroplasty. Study subjects (n = 29) underwent consecutive, standardized, plain radiographic examinations of the hip on the same day after repositioning. The resulting radiograph pairs were digitized and analyzed using EBRA. The precision (95% confidence interval) of the method for measuring migration and wear was < +/-0.9 mm for both implant components. The 95% confidence intervals for measurement of cup inclination and anteversion and femoral stem/shaft angle were < +/-1.7degrees. Measurement precision was not strongly related to patient gender, digitization method, or observer. The EBRA-Digital method has sufficient precision to detect clinically relevant migration to allow individual patient monitoring after total hip arthroplasty. The method requires careful patient positioning and radiographic technique to produce consistently images suitable for analysis.
引用
收藏
页码:910 / 916
页数:7
相关论文
共 15 条
[1]
Tantalum ball position after total hip arthroplasty [J].
Eldridge, JDJ ;
Avramidis, K ;
Lee, M ;
Learmonth, ID .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (03) :414-416
[2]
EARLY MIGRATION AND LATE ASEPTIC FAILURE OF PROXIMAL FEMORAL PROSTHESES [J].
FREEMAN, MAR ;
PLANTEBORDENEUVE, P .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (03) :432-438
[3]
Migration of the uncemented Harris-Galante acetabular cup - Results of the einbildroentgenanalyse (EBRA) method [J].
Hendrich, C ;
Bahlmann, J ;
Eulert, J .
JOURNAL OF ARTHROPLASTY, 1997, 12 (08) :889-895
[4]
Iichmann T, 1998, ACTA ORTHOP SCAND, V69, P119
[5]
MEASUREMENT ACCURACY IN ACETABULAR CUP WEAR - 3 RETROSPECTIVE METHODS COMPARED WITH ROENTGEN STEREOPHOTOGRAMMETRY [J].
ILCHMANN, T ;
MJOBERG, B ;
WINGSTRAND, H .
JOURNAL OF ARTHROPLASTY, 1995, 10 (05) :636-642
[6]
Migration and wear of long-term successful Charnley total hip replacements [J].
Ilchmann, T ;
Markovic, L ;
Joshi, A ;
Hardinge, K ;
Murphy, J ;
Wingstrand, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (03) :377-381
[7]
Ilchmann T, 1992, J Arthroplasty, V7, P121, DOI 10.1016/0883-5403(92)90004-A
[8]
DOES EARLY MICROMOTION OF FEMORAL STEM PROSTHESES MATTER - 4-7-YEAR STEREORADIOGRAPHIC FOLLOW-UP OF 84 CEMENTED PROSTHESES [J].
KARRHOLM, J ;
BORSSEN, B ;
LOWENHIELM, G ;
SNORRASON, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06) :912-917
[9]
Karrholm J, 1997, CLIN ORTHOP RELAT R, P94
[10]
Early radiological observations may predict the long-term survival of femoral hip prostheses [J].
Kobayashi, A ;
Donnelly, WJ ;
Scott, G ;
Freeman, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (04) :583-589