Pneumatic impedance control of a 3-d.o.f. physiotherapy robot

被引:12
作者
Richardson, R. [1 ]
Jackson, A.
Culmer, P.
Bhakta, B.
Levesley, M. C.
机构
[1] Univ Manchester, Sch Comp Sci, Manchester M13 9PL, Lancs, England
[2] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[3] Univ Leeds, Sch Rheumatol, Leeds LS2 9JT, W Yorkshire, England
[4] Univ Leeds, Sch Med, Rehabil Res Unit, Leeds LS2 9JT, W Yorkshire, England
关键词
impedance control; pneumatic cylinders; robotic physiotherapy;
D O I
10.1163/156855306778960590
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Stroke is a common condition resulting in 30 000 people per annum left with significant disability. In patients with severe arm paresis after stroke, functional recovery in the affected arm is poor. Inadequate intensity of treatment is cited as one factor accounting for the lack arm recovery found in some studies. Given that physical therapy resource is limited, strategies to enhance the physiotherapists' efforts are needed. One approach is to use robotic techniques to augment movement therapy. A 3-d.o.f. pneumatic robot has been developed to apply physiotherapy to the upper limb. The robot has been designed with a workspace encompassing the reach-retrieve range of the average male. Slight non-linearities in the response of the pneumatic system are observed and explained. Building upon previous work that used an error-prone custom force sensor, a commercial 6-d.o.f. force sensor is used to apply impedance control in 3 d.o.f. on the robot.
引用
收藏
页码:1321 / 1339
页数:19
相关论文
共 27 条
[1]   A PROSPECTIVE-STUDY OF ACUTE CEREBROVASCULAR-DISEASE IN THE COMMUNITY - THE OXFORDSHIRE-COMMUNITY-STROKE-PROJECT 1981-86 .1. METHODOLOGY, DEMOGRAPHY AND INCIDENT CASES OF 1ST-EVER STROKE [J].
BAMFORD, J ;
SANDERCOCK, P ;
DENNIS, M ;
WARLOW, C ;
JONES, L ;
MCPHERSON, K ;
VESSEY, M ;
FOWLER, G ;
MOLYNEUX, A ;
HUGHES, T ;
BURN, J ;
WADE, D .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1988, 51 (11) :1373-1380
[2]  
BUCKLEY MA, MULOS SYSTEMS PROTOT
[3]   THE PARALLEL APPROACH TO FORCE POSITION CONTROL OF ROBOTIC MANIPULATORS [J].
CHIAVERINI, S ;
SCIAVICCO, L .
IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 1993, 9 (04) :361-373
[4]  
Cozens J A, 1999, IEEE Trans Rehabil Eng, V7, P254, DOI 10.1109/86.769416
[5]   Joint impedance pneumatic control for multilink systems [J].
Gorce, P ;
Guihard, M .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 1999, 121 (02) :293-297
[6]   Position-based impedance control of an industrial hydraulic manipulator [J].
Heinrichs, B ;
Sepehri, N ;
ThorntonTrump, AB .
IEEE CONTROL SYSTEMS MAGAZINE, 1997, 17 (01) :46-52
[7]   IMPEDANCE CONTROL - AN APPROACH TO MANIPULATION .1. THEORY [J].
HOGAN, N .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 1985, 107 (01) :1-7
[8]  
Krebs H I, 1998, IEEE Trans Rehabil Eng, V6, P75, DOI 10.1109/86.662623
[9]   Intensity of leg and arm training after primary middle-cerebral-artery stroke: a randomised trial [J].
Kwakkel, G ;
Wagenaar, RC ;
Twisk, JWR ;
Lankhorst, GJ ;
Koetsier, JC .
LANCET, 1999, 354 (9174) :191-196
[10]   AN ANALYSIS OF A PNEUMATIC SERVO SYSTEM AND ITS APPLICATION TO A COMPUTER-CONTROLLED ROBOT [J].
LIU, S ;
BOBROW, JE .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 1988, 110 (03) :228-235