MR Elastography as a Method for the Assessment of Myocardial Stiffness: Comparison with an Established Pressure-Volume Model in a Left Ventricular Model of the Heart

被引:63
作者
Kolipaka, Arunark [1 ]
McGee, Kiaran P. [1 ]
Araoz, Philip A. [1 ]
Glaser, Kevin J. [1 ]
Manduca, Armando [1 ]
Romano, Anthony J. [2 ]
Ehman, Richard L. [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
[2] USN, Res Lab, Washington, DC 20375 USA
基金
美国国家卫生研究院;
关键词
MRE; P-V relationships; gated MRE sequences; cardiac MRE; MAGNETIC-RESONANCE ELASTOGRAPHY; FAILURE; MUSCLE; ELASTICITY; MOTION; SENC;
D O I
10.1002/mrm.21991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance elastography (MRE) measurements of shear stiffness (mu) in a spherical phantom experiencing both static and cyclic pressure variations were compared to those derived from an established pressure-volume (P-V)-based model. A spherical phantom was constructed using a silicone rubber composite of 10 cm inner diameter and 1.3 cm thickness. A gradient echo MRE sequence was used to determine mu within the phantom at static and cyclic pressures ranging from 55 to 90 mmHg. Average values of mu using MRE were obtained within a region of interest and were compared to the P-V-derived estimates. Under both static and cyclic pressure conditions, the P-V- and MRE-based estimates of mu ranged from 98.2 to 155.1 kPa and 96.2 to 150.8 kPa, respectively. Correlation coefficients (R-2) of 0.98 and 0.97 between the P-V and MRE-based estimates of shear stiffness measurements were obtained. For both static and cyclic pressures, MRE-based measures of mu agree with those derived from a P-V model, suggesting that MRE can be used as a new, noninvasive method of assessing mu in sphere-like fluid-filled organs such as the heart. Magn Reson Med 62:135-140, 2009. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:135 / 140
页数:6
相关论文
共 28 条
[1]   Determination of thigh muscle stiffness using magnetic resonance elastography [J].
Bensamoun, SF ;
Ringleb, SI ;
Littrell, L ;
Chen, QS ;
Brennan, M ;
Ehman, RL ;
An, KN .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 23 (02) :242-247
[2]   Assessment of systolic and diastolic ventricular properties via pressure-volume analysis: a guide for clinical, translational, and basic researchers [J].
Burkhoff, D ;
Mirsky, I ;
Suga, H .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2005, 289 (02) :H501-H512
[3]  
Dresner MA, 2001, J MAGN RESON IMAGING, V13, P269, DOI 10.1002/1522-2586(200102)13:2<269::AID-JMRI1039>3.0.CO
[4]  
2-1
[5]  
Fowlkes J. B., 1995, Biophysical Bases of Elasticity Imaging, P223, DOI DOI 10.1007/978-1-4615-1943-0_23
[6]   Real-time MR imaging of myocardial regional function using strain-encoding (SENC) with tissue through-plane motion tracking [J].
Ibrahim, El-Sayed H. ;
Stuber, Matthias ;
Fahmy, Ahmed S. ;
Abd-Elmoniem, Khaled Z. ;
Sasano, Tetsuo ;
Abraham, M. Roselle ;
Osman, Nael F. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (06) :1461-1470
[7]  
JUNGER CM, 1972, SOUND STRUCTURES THE, P230
[8]  
KOLIPAKA A, 2008, P 16 ANN M ISMRM TOR, P1030
[9]   Heart failure - Current perspectives in diastolic dysfunction and diastolic heart failure [J].
Leite-Moreira, AF .
HEART, 2006, 92 (05) :712-718
[10]   Magnetic resonance elastography: Non-invasive mapping of tissue elasticity [J].
Manduca, A ;
Oliphant, TE ;
Dresner, MA ;
Mahowald, JL ;
Kruse, SA ;
Amromin, E ;
Felmlee, JP ;
Greenleaf, JF ;
Ehman, RL .
MEDICAL IMAGE ANALYSIS, 2001, 5 (04) :237-254