GATED IRRADIATION WITH SCANNED PARTICLE BEAMS

被引:69
作者
Bert, Christoph [1 ]
Gemmel, Alexander [1 ]
Saito, Nami [1 ]
Rietzel, Eike [1 ,2 ]
机构
[1] Gesell Schwerionenforsch mbH, Biophys Abt, D-64291 Darmstadt, Germany
[2] Siemens Healthcare, Particle Therapy, Erlangen, Germany
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2009年 / 73卷 / 04期
关键词
Particle therapy; Gating; Scanning; Target motion; HEAVY-ION RADIOTHERAPY; DOSE OPTIMIZATION; SYSTEM; VERIFICATION; THERAPY; DESIGN; MOTION; FIELDS; TUMORS;
D O I
10.1016/j.ijrobp.2008.11.014
中图分类号
R73 [肿瘤学];
学科分类号
100214 [肿瘤学];
摘要
Purpose: To demonstrate mitigation of the interplay effects of scanned particle beams and residual target motion within a gating window by increased overlap of pencil beams. Methods and Materials: Lateral overlap was increased by increasing the pencil beam widths or by decreasing the distance between the pencil beams (scan grid). Longitudinal overlap was increased by reducing the distance between iso-range slices. For scanned carbon ion beams, simulation studies were performed and validated experimentally to determine the required parameters for different residual motion characteristics. The dose distributions were characterized by the maximal local deviations representing local over- and underdosage. Results: For residual lateral motion, the local deviations were <5% for 2, 4, and 7 mm residual motion within the gating window for a 2-mm scan grid and pencil beams of 10, 14, and 18 mm full width at half maximum, respectively. Decreasing the iso-range slice distance from 3 mm to 1 mm effectively mitigated <= 10 mm water-equivalent range changes. Experimental data reproduced the simulation results. Conclusion: In charged particle therapy with a scanned beam, interplay effects between gated beam delivery and residual target motion can be decreased effectively by increasing the overlap between pencil beams laterally, as well as longitudinally. (C) 2009 Elsevier Inc.
引用
收藏
页码:1270 / 1275
页数:6
相关论文
共 21 条
[1]
Quantification of interplay effects of scanned particle beams and moving targets [J].
Bert, Christoph ;
Groezinger, Sven O. ;
Rietzel, Eike .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2253-2265
[2]
4D treatment planning for scanned ion beams [J].
Bert, Christoph ;
Rietzel, Eike .
RADIATION ONCOLOGY, 2007, 2 (1)
[3]
Target motion tracking with a scanned particle beam [J].
Bert, Christoph ;
Saito, Nami ;
Schmidt, Alexander ;
Chaudhri, Naved ;
Schardt, Dieter ;
Rietzel, Eike .
MEDICAL PHYSICS, 2007, 34 (12) :4768-4771
[4]
Four-dimensional proton treatment planning for lung tumors [J].
Engelsman, M ;
Rietzel, E ;
Kooy, HM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (05) :1589-1595
[5]
Design study of a raster scanning system for moving target irradiation in heavy-ion radiotherapy [J].
Furukawa, Takuji ;
Inaniwa, Taku ;
Sato, Shinji ;
Tomitani, Takehiro ;
Minohara, Shinichi ;
Noda, Koji ;
Kanai, Tatsuaki .
MEDICAL PHYSICS, 2007, 34 (03) :1085-1097
[6]
Biological dose optimization with multiple ion fields [J].
Gemmel, A. ;
Hasch, B. ;
Ellerbrock, M. ;
Weyrather, W. K. ;
Kraemer, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23) :6991-7012
[7]
MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY [J].
HABERER, T ;
BECHER, W ;
SCHARDT, D ;
KRAFT, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) :296-305
[8]
JAKEL O, 2008, COMMISSIONING HEIDEL
[9]
A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ions [J].
Karger, CP ;
Jäkel, O ;
Hartmann, GH ;
Heeg, P .
MEDICAL PHYSICS, 1999, 26 (10) :2125-2132
[10]
Treatment planning for heavy-ion radiotherapy:: physical beam model and dose optimization [J].
Krämer, M ;
Jäkel, O ;
Haberer, T ;
Kraft, G ;
Schardt, D ;
Weber, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3299-3317