Design of a capacitively decoupled transmit/receive NMR phased array for high field microscopy at 14.1 T

被引:56
作者
Zhang, XZ
Webb, A [1 ]
机构
[1] Univ Illinois, Beckman Inst Adv Sci & Technol, Dept Elect & Comp Engn, Urbana, IL 61801 USA
[2] Univ Wurzburg, Inst Phys, D-8700 Wurzburg, Germany
关键词
D O I
10.1016/j.jmr.2004.05.004
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A design is presented for a "phased array" of four transmit/receive saddle-geometry volume coils for microimaging at 600 MHz within a 45 mm clear-bore vertical magnet. The small size of the coils, similar to10 mm in length, and high frequency of operation both present considerable challenges for the design of a phased array. The particular design consists of four saddle coils, stacked vertically, in order to produce an array suitable for imaging samples, typical of many microimaging studies, with a large length:diameter ratio. Optimal coil overlap is used to reduce the mutual inductance between adjacent coils, and capacitive networks are used to maximize the isolation between all of the coils. Standard 50 Omega input impedance preamplifiers are used so that the preamplifiers do not have to be integrated directly into the probe. Isolation between coils was better than 20 dB for all coil pairs. An increase in signal-to-noise of 70 +/- 3% was achieved, averaged over the whole array, compared to a single coil of the same dimensions. High resolution phased array images are shown for ex vivo tissue samples. (C) 2004 Published by Elsevier Inc.
引用
收藏
页码:149 / 155
页数:7
相关论文
共 34 条
[1]  
Adriany G., 2003, P 11 ANN M ISMRM TOR, P474
[2]  
Bankson JA, 2000, MAGN RESON MATER PHY, V10, P93
[3]  
Barberi EA, 2000, MAGNET RESON MED, V43, P284, DOI 10.1002/(SICI)1522-2594(200002)43:2<284::AID-MRM16>3.0.CO
[4]  
2-C
[5]  
Barfuss H, 1990, NMR Biomed, V3, P31, DOI 10.1002/nbm.1940030106
[6]   Progress in high field MRI at the University of Florida [J].
Beck, B ;
Plant, DH ;
Grant, SC ;
Thelwall, PE ;
Silver, X ;
Mareci, TH ;
Benveniste, H ;
Smith, M ;
Collins, C ;
Crozier, S ;
Blackband, SJ .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2002, 13 (03) :152-157
[7]   Phased array imaging on a 4.7T/33cm animal research system [J].
Beck, BL ;
Blackband, SJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (11) :4292-4294
[8]   What is the optimum phased array coil design for cardiac and torso magnetic resonance? [J].
Bottomley, PA ;
Olivieri, CHL ;
Giaquinto, R .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (04) :591-599
[9]   Design of a SENSE-optimized high-sensitivity MRI receive coil for brain imaging [J].
de Zwart, JA ;
Ledden, PJ ;
Kellman, P ;
van Gelderen, P ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1218-1227
[10]  
Fox T.R., 1989, P 8 ANN M SOC MAGN R, P99