IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119

被引:790
作者
Ezzell, Gary A. [2 ]
Burmeister, Jay W. [3 ]
Dogan, Nesrin [4 ]
LoSasso, Thomas J. [5 ]
Mechalakos, James G. [5 ]
Mihailidis, Dimitris [6 ]
Molineu, Andrea [7 ]
Palta, Jatinder R. [8 ]
Ramsey, Chester R. [9 ]
Salter, Bill J. [10 ]
Shi, Jie [11 ]
Xia, Ping [12 ]
Yue, Ning J. [13 ]
Xiao, Ying [1 ]
机构
[1] Thomas Jefferson Univ Hosp, Dept Radiat Oncol, Philadelphia, PA 19107 USA
[2] Mayo Clin Scottsdale, Dept Radiat Oncol, MCSB Concourse, Phoenix, AZ 85054 USA
[3] Wayne State Univ, Sch Med, Karmanos Canc Ctr, Detroit, MI 48201 USA
[4] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
[5] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10065 USA
[6] Charleston Radiat Therapy Cons, Dept Radiat Oncol & Med Phys, Charleston, WV 25304 USA
[7] Univ Texas MD Anderson Canc Ctr, RPC, Houston, TX 77030 USA
[8] Univ Florida, Hlth Sci Ctr, Dept Radiat Oncol, Gainesville, FL 32610 USA
[9] Thompson Canc Survival Ctr, Knoxville, TN 37916 USA
[10] Univ Utah, Dept Radiat Oncol, Salt Lake City, UT 84112 USA
[11] Sun Nucl Corp, Melbourne, FL 32940 USA
[12] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
[13] Canc Inst New Jersey, Dept Radiat Oncol, New Brunswick, NJ 08901 USA
关键词
IMRT; commissioning; quality assurance; MODULATED RADIATION-THERAPY; DOSE UNCERTAINTY MODEL; QUALITY-ASSURANCE; DELIVERY; IMPLEMENTATION; TOLERANCES; ACCURACY;
D O I
10.1118/1.3238104
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit=vertical bar mean vertical bar + 1.96 sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3238104]
引用
收藏
页码:5359 / 5373
页数:15
相关论文
共 23 条
[1]
Alber M., 2008, Guidelines for the verification of IMRT
[2]
Both S, 2007, J APPL CLIN MED PHYS, V8, P1
[3]
Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system [J].
Cadman, P ;
Bassalow, R ;
Sidhu, NPS ;
Ibbott, G ;
Nelson, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :3001-3010
[4]
Intensity-modulated radiation therapy dose prescription, recording, and delivery: Patterns of variability among institutions and treatment planning systems [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Chopra, Kashmiri L. ;
Mitra, Raj K. ;
Srivastava, Shiv P. ;
Glatstein, Eli .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2008, 100 (05) :300-307
[5]
Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[6]
American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[7]
An inter-centre quality assurance network for IMRT verification: Results of the ESTRO QUASIMODO project [J].
Gillis, S ;
De Wagter, C ;
Bohsung, J ;
Perrin, B ;
Williams, P ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (03) :340-353
[8]
Special Dosimetric/Measurement considerations in commissioning a novel integrated MiniMLC linear accelerator [J].
Harrison, A. ;
Misic, V. ;
Podder, T. ;
Bednarz, G. ;
Cryan, G. ;
Fallon, K. ;
Houser, C. ;
Xiao, Y. .
MEDICAL PHYSICS, 2007, 34 (06) :2489-2490
[9]
Challenges in Credentialing Institutions and Participants in Advanced Technology Multi-institutional Clinical Trials [J].
Ibbott, Geoffrey S. ;
Followill, David S. ;
Molineu, H. Andrea ;
Lowenstein, Jessica R. ;
Alvarez, Paola E. ;
Roll, Joye E. .
International Journal of Radiation Oncology Biology Physics, 2008, 71 (1 SUPPL.) :S71-S75
[10]
A generalized a priori dose uncertainty model of IMRT delivery [J].
Jin, Hosang ;
Palta, Jatinder ;
Suh, Tae-Suk ;
Kim, Siyong .
MEDICAL PHYSICS, 2008, 35 (03) :982-996