Dosimetric properties of the Theraview™ fluoroscopic electronic portal imaging device

被引:21
作者
Glendinning, AG [1 ]
Bonnett, DE [1 ]
机构
[1] Leicester Royal Infirm NHS Trust, Dept Phys Med, Leicester LE1 5WW, Leics, England
关键词
D O I
10.1259/bjr.73.869.10884749
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Electronic portal imaging devices (EPIDs) can be used for non-imaging applications in radiotherapy such as patient dosimetry. Of the systems available, the fluoroscopic camera-based EPID Theraview(TM) (InfiMed Inc.) has not been studied to date, and a review of the dosimetric properties of the system is presented here. In the "single set-up" mode of image acquisition, pixel intensity increases sublinearly with applied dose. The response was dependent on the system's video signal gain and showed a threshold dose to the detector in the range 0.05-0.35 cGy, and pixel saturation at detector doses in the range 1.2-1.6 cGy. Repeated exposures of the EPID were observed to be extremely reproducible (standard deviation 0.5%). The sensitivity of the system showed a linear decline of 0.04% day(-1) over a 68-day period, during which time the relative off-axis response within a 10 x 10 cm(2) field was constant to within a standard deviation of 0.56%. The system shows spatial non-uniformity, which requires correction for application to dose measurements in two-dimensions. Warm-up of the camera control unit required a period of at least 40 min and was associated with an enhancement in pixel intensity of up to 12%. A radiation dose history effect was observed at doses as low as 0.2 Gy. Camera dark current was shown to be negligible at normal accelerator operation. No discernible image distortion was found. Mechanical stability on gantry rotation was also assessed and image displacement of up to 5 mm at the isocentre was observed. It was concluded that the device could be used for dosimetry provided necessary precautions were observed and corrections made.
引用
收藏
页码:517 / 530
页数:14
相关论文
共 48 条
[1]   CHARACTERIZATION OF FLUORESCENT SCREENS FOR IMAGING APPLICATIONS WITH MEV NEUTRONS AND PHOTONS [J].
BERZINS, GJ ;
LUMPKIN, AH ;
SMITH, HL .
OPTICAL ENGINEERING, 1983, 22 (05) :633-642
[2]   Fast 2D phantom dosimetry for scanning proton beams [J].
Boon, SN ;
van Luijk, P ;
Schippers, JM ;
Meertens, H ;
Denis, JM ;
Vynckier, S ;
Medin, J ;
Grusell, E .
MEDICAL PHYSICS, 1998, 25 (04) :464-475
[3]  
Conte L, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P279
[4]  
Curtin-Savard A, 1997, Med Dosim, V22, P101, DOI 10.1016/S0958-3947(97)00002-2
[5]   Verification of segmented beam delivery using a commercial electronic portal imaging device [J].
Curtin-Savard, AJ ;
Podgorsak, EB .
MEDICAL PHYSICS, 1999, 26 (05) :737-742
[6]  
Deych R, 1995, P INT C IN SCINT THE, P36
[7]   DAILY DOSIMETRIC QUALITY-CONTROL OF THE MM50 RACETRACK MICROTRON USING AN ELECTRONIC PORTAL IMAGING DEVICE [J].
DIRKX, MLP ;
KROONWIJK, M ;
DEBOER, JCJ ;
HEIJMEN, BJM .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :55-60
[8]  
DORNER KJ, 1995, MED PHYS 95 RONTG GE, P260
[9]   Dosimetric characteristics of a liquid-filled electronic portal imaging device [J].
Essers, M ;
Hoogervorst, BR ;
vanHerk, M ;
Lanson, H ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1265-1272
[10]   Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multi-level scheme algebraic reconstruction technique (MLS-ART) [J].
Guan, HQ ;
Zhu, YP .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2925-2937