Single camera based spectral domain polarization sensitive optical coherence tomography

被引:68
作者
Baumann, Bernhard [1 ]
Goetzinger, Erich [1 ]
Pircher, Michael [1 ]
Hitzenberger, Christoph K. [1 ]
机构
[1] Med Univ vienna, Ctr Biomed Eng & Phys, A-1090 Vienna, Austria
来源
OPTICS EXPRESS | 2007年 / 15卷 / 03期
基金
奥地利科学基金会;
关键词
D O I
10.1364/OE.15.001054
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We developed a new spectral domain polarization sensitive optical coherence tomography (SD PS-OCT) system that requires only a single spectrometer CCD camera. The spectra of the horizontal and vertical polarization channels are imaged adjacent to each other on a 2048 pixel line scan camera, using 1024 pixels for each channel. Advantages of the system are reduced costs and complexity, lower demands on timing and triggering circuitry, and higher robustness against camera misalignments. We discuss the additional postprocessing required to accommodate for spectral distortions, show calibration measurements in a test sample, and finally demonstrate the system for measuring human ocular tissue in vivo. (c) 2007 Optical Society of America.
引用
收藏
页码:1054 / 1063
页数:10
相关论文
共 31 条
[1]   Polarization-sensitive optical coherence tomography of dental structures [J].
Baumgartner, A ;
Dichtl, S ;
Hitzenberger, CK ;
Sattmann, H ;
Robl, B ;
Moritz, A ;
Fercher, ZF ;
Sperr, W .
CARIES RESEARCH, 2000, 34 (01) :59-69
[2]  
Bouma B.E., 2002, Handbook of Optical Coherence Tomography
[3]   Spectral-domain polarization-sensitive optical coherence tomography at 850 nm [J].
Cense, B ;
Chen, TC ;
Mujat, M ;
Joo, C ;
Akkin, T ;
Park, BH ;
Pierce, MC ;
Yun, A ;
Bouma, BE ;
Teamey, GJ ;
de Boer, JF .
Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine IX, 2005, 5690 :159-162
[4]   In vivo depth-resolved birefringence measurements of the human retinal nerve fiber layer by polarization-sensitive optical coherence tomography [J].
Cense, B ;
Chen, TC ;
Park, BH ;
Pierce, MC ;
de Boer, JF .
OPTICS LETTERS, 2002, 27 (18) :1610-1612
[5]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[6]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[7]   Determination of the depth-resolved Stokes parameters of light backscattered from turbid media by use of polarization-sensitive optical coherence tomography [J].
de Boer, JF ;
Milner, TE ;
Nelson, JS .
OPTICS LETTERS, 1999, 24 (05) :300-302
[8]   Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography [J].
deBoer, JF ;
Milner, TE ;
vanGemert, MJC ;
Nelson, JS .
OPTICS LETTERS, 1997, 22 (12) :934-936
[9]   Spectral resolution and sampling issues in Fourier-transform spectral interferometry [J].
Dorrer, C ;
Belabas, N ;
Likforman, JP ;
Joffre, M .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2000, 17 (10) :1795-1802
[10]   MEASUREMENT OF INTRAOCULAR DISTANCES BY BACKSCATTERING SPECTRAL INTERFEROMETRY [J].
FERCHER, AF ;
HITZENBERGER, CK ;
KAMP, G ;
ELZAIAT, SY .
OPTICS COMMUNICATIONS, 1995, 117 (1-2) :43-48