Quantitative imaging and correction for cascade gamma radiation of 76Br with 2D and 3D PET

被引:55
作者
Lubberink, M [1 ]
Schneider, H
Bergström, M
Lundqvist, H
机构
[1] Uppsala Univ, Rudbeck Lab, Sect Biomed Radiat Sci, S-75185 Uppsala, Sweden
[2] Uppsala Univ, PET Ctr, Univ Hosp, Uppsala Res Imaging Solut AB, S-75185 Uppsala, Sweden
关键词
D O I
10.1088/0031-9155/47/19/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Several positron emitting nuclides with applications in PET, such as Br-76, I-124, In-110m and Y-86, also emit gamma radiation in their decays. Measured coincidences between annihilation photons and this cascade gamma radiation are essentially true coincidences and the standard PET corrections do not account for them. We investigated the performance of Br-76 in 2D and 3D PET, the effect of the gamma radiation emitted by Br-76 on quantitative accuracy and the distribution of cascade gamma radiation coincidences in 2D and 3D PET sinograms. A correction method for cascade gamma radiation coincidences was implemented and evaluated. Count rate linearity was affected by the gamma radiation from the Br-76 decay. Spatial resolution and sphere recovery were slightly worse for Br-76 compared to F-18. Correction for cascade gamma radiation coincidences by subtraction of a linear projection tail fit improved total correction accuracy to similar values as for positron-only emitters such as F-18, and improved image contrast significantly.
引用
收藏
页码:3519 / 3534
页数:16
相关论文
共 26 条
[11]  
HERZOG H, 1993, J NUCL MED, V34, P2222
[12]  
KARP JS, 1991, J NUCL MED, V32, P2342
[13]  
KOHLMYER SG, 1999, J NUCL MED, V40, P280
[14]   PERFORMANCE-CHARACTERISTICS OF AN 8-RING WHOLE-BODY PET SCANNER [J].
KOPS, ER ;
HERZOG, H ;
SCHMID, A ;
HOLTE, S ;
FEINENDEGEN, LE .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (03) :437-445
[15]  
Lövqvist A, 1999, MED PHYS, V26, P249, DOI 10.1118/1.598512
[16]  
LUBBERINK M, 1999, FUTURE DIRECTIONS NU
[17]  
LUBBERINK M, 2002, J NUCL MED, V43
[18]  
Lundqvist H, 1999, ACTA ONCOL, V38, P335
[19]   Quantitative PET with positron emitters that emit prompt gamma rays [J].
Martin, CC ;
Christian, BT ;
Satter, MR ;
Nickerson, LDH ;
Nickles, RJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (04) :681-687
[20]  
Pentlow, 2000, Clin Positron Imaging, V3, P85, DOI 10.1016/S1095-0397(00)00046-7