Inappropriate pauses during bradycardia pacing in a third-generation implantable cardioverter defibrillator

被引:5
作者
Cossu, SF [1 ]
Hsia, HH [1 ]
Simson, MB [1 ]
Hanna, MS [1 ]
Clyne, CA [1 ]
机构
[1] UNIV PENN,SCH MED,DEPT MED,DIV CARDIOVASC,PHILADELPHIA,PA 19104
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 09期
关键词
implantable cardioverter defibrillators; pacing; sensing algorithms;
D O I
10.1111/j.1540-8159.1997.tb04248.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many ICD devices have the capability for back up bradycardia pacing. Because of the use of a single sensing algorithm for both bradycardic and tachycardia functions, they may be prone to certain ''sensing errors.'' Following implantation of an ICD in a patient with long QT syndrome, ''inappropriate'' pauses were noted during bradycardia pacing, which were exactly twice the programmed pacing cycle length. This was due to an automatic increase in the device's sensitivity during pacing, a characteristic of the automatic gain control of this particular ICD. Proper recognition of this ICD's special features, known as ''lower threshold crossing,'' allowed noninvasive rectification of the problem and prevented these pauses.
引用
收藏
页码:2271 / 2274
页数:4
相关论文
共 13 条
[1]   FAILURE OF THE AUTOMATIC IMPLANTABLE DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION [J].
BARDY, GH ;
IVEY, TD ;
STEWART, R ;
GRAHAM, EL ;
GREENE, HL .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 58 (11) :1107-1108
[2]   CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALKINS, H ;
BRINKER, J ;
VELTRI, EP ;
GUARNIERI, T ;
LEVINE, JH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :666-673
[3]   PACED BEATS FOLLOWING SINGLE NONSENSED COMPLEXES IN A CODEPENDENT CARDIOVERTER DEFIBRILLATOR AND BRADYCARDIA PACING SYSTEM - POTENTIAL FOR VENTRICULAR-TACHYCARDIA INDUCTION [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08) :1281-1287
[4]   UNIQUE SENSING ERRORS IN 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALLANS, DJ ;
HOOK, BG ;
KLEIMAN, RB ;
MITRA, RL ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1135-1140
[5]   CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
ECHT, DS ;
ARMSTRONG, K ;
SCHMIDT, P ;
OYER, PE ;
STINSON, EB ;
WINKLE, RA .
CIRCULATION, 1985, 71 (02) :289-296
[6]   IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - VENTRITEX(R) CADENCE(R) [J].
FAIN, ES ;
WINKLE, RA .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1993, 4 (02) :211-223
[7]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS AND SURVIVAL IN PATIENTS WITH MALIGNANT VENTRICULAR ARRHYTHMIAS [J].
KELLY, PA ;
CANNOM, DS ;
GARAN, H ;
MIRABAL, GS ;
HARTHORNE, JW ;
HURVITZ, RJ ;
VLAHAKES, GJ ;
JACOBS, ML ;
ILVENTO, JP ;
BUCKLEY, MJ ;
RUSKIN, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1278-1286
[8]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS, AND DEVICE FAILURES [J].
MARCHLINSKI, FE ;
FLORES, BT ;
BUXTON, AE ;
HARGROVE, WC ;
ADDONIZIO, VP ;
STEPHENSON, LW ;
HARKEN, AH ;
DOHERTY, JU ;
GROGAN, EW ;
JOSEPHSON, ME .
ANNALS OF INTERNAL MEDICINE, 1986, 104 (04) :481-488
[9]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - T-WAVE SENSING IN THE NEWEST GENERATION [J].
SINGER, I ;
DEBORDE, R ;
VELTRI, EP ;
SIDDOWAY, LA ;
LEVINE, JH ;
GRIFFITH, LSC ;
GUARNIERI, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1584-1591
[10]   POTENTIAL HAZARDS OF FIXED GAIN SENSING AND ARRHYTHMIA RECONFIRMATION FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
SINGER, I ;
ADAMS, L ;
AUSTIN, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (05) :1070-1084