On-line preconcentration and determination of copper, lead and chromium(VI) using unloaded polyurethane foam packed column by flame atomic absorption spectrometry in natural waters and biological samples

被引:141
作者
Anthemidis, AN [1 ]
Zachariadis, GA [1 ]
Stratis, JA [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Chem, Analyt Chem Lab, GR-54006 Thessaloniki, Greece
关键词
atomic absorption spectrometry; on-line preconcentration; polyurethane foam; copper; lead; chromium;
D O I
10.1016/S0039-9140(02)00373-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, sensitive and low cost, flow injection time-based method was developed for on-line preconcentration and determination of copper, lead and chromium(VI) at sub mug l(-1) levels in natural waters and biological samples. At the optimum pH, the on-line formed metal-ammonium pyrrolidine dithiocarbamate (APDC) complexes were sorbed on the unloaded commercial polyurethane foam (PUF), and subsequent eluted quantitatively by isobutylmethylketone and determined by flame atomic absorption spectrometry (FAAS). All chemical, and flow injection variables were optimized for the quantitative preconcentration of each metal and a study of interference level of various ions was also carried out. The system offered improved flexibility, low backpressure and applicability to all the studied metals. At a sample frequency of 36 h(-1) and a 60 s preconcentration time, the enhancement factor was 170, 131 and 28, the detection limit was 0.2, 1.8 and 2.0 mug l(-1) and the precision, expressed as relative standard deviation (s(r)), was 2.8 (at 10 mug l(-1)), 3.4 (at 50 mug l(-1)) and 3.6% (at 50 mug l(-1)) for Cu(II), Pb(II) and Cr(VI), respectively. The accuracy of the developed method was sufficient and evaluated by the analysis of certified reference materials and spiked water samples. Finally, the method was applied to the analysis of environmental samples. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:831 / 840
页数:10
相关论文
共 46 条
[21]   FLAME ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF LEAD IN BIOLOGICAL SAMPLES USING A FLOW-INJECTION SYSTEM WITH ONLINE PRECONCENTRATION BY COPRECIPITATION WITHOUT FILTRATION [J].
FANG, ZL ;
SPERLING, M ;
WELZ, B .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (04) :301-306
[22]   DETERMINATION OF CADMIUM, LEAD AND COPPER IN WATER SAMPLES BY FLAME ATOMIC-ABSORPTION SPECTROMETRY WITH PRECONCENTRATION BY FLOW-INJECTION ONLINE SORBENT EXTRACTION [J].
FANG, ZL ;
GUO, TZ ;
WELZ, B .
TALANTA, 1991, 38 (06) :613-619
[23]   An on-line continuous flow system for copper enrichment and determination by flame atomic absorption spectroscopy [J].
Ferreira, SLC ;
Lemos, VA ;
Moreira, BC ;
Costa, ACS ;
Santelli, RE .
ANALYTICA CHIMICA ACTA, 2000, 403 (1-2) :259-264
[24]   An automated on-line flow system for the pre-concentration and determination of lead by flame atomic absorption spectrometry [J].
Ferreira, SLC ;
Lemos, VA ;
Santelli, RE ;
Ganzarolli, E ;
Curtius, AJ .
MICROCHEMICAL JOURNAL, 2001, 68 (01) :41-46
[25]   An on-line solid phase extraction system using polyurethane foam for the spectrophotometric determination of nickel in silicates and alloys [J].
Ferreira, SLC ;
de Jesus, DS ;
Cassella, RJ ;
Costa, ACS ;
de Carvalho, MS ;
Santelli, RE .
ANALYTICA CHIMICA ACTA, 1999, 378 (1-3) :287-292
[26]   FULLERENES AS SORBENT MATERIALS FOR METAL PRECONCENTRATION [J].
GALLEGO, M ;
DEPENA, YP ;
VALCARCEL, M .
ANALYTICAL CHEMISTRY, 1994, 66 (22) :4074-4078
[27]   Possibilities for the simultaneous preconcentration and flame atomic absorption spectrometric determination of Cr(III) and Cr(VI) using a C18 column and sorption loop [J].
Gáspár, A ;
Sógor, C ;
Posta, J .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 363 (5-6) :480-483
[28]   On-line sorption preconcentration of chromium(VI) and its determination by flame atomic absorption spectrometry [J].
Gaspar, A ;
Posta, J .
ANALYTICA CHIMICA ACTA, 1997, 354 (1-3) :151-158
[29]   TRACE-METAL ENRICHMENT BY AUTOMATED ONLINE COLUMN PRECONCENTRATION FOR FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
HIRATA, S ;
HONDA, K ;
KUMAMARU, T .
ANALYTICA CHIMICA ACTA, 1989, 221 (01) :65-76
[30]  
Kartikeyan S, 1997, ANAL LETT, V30, P1037