Magnetic resonance image-guided transcatheter closure of atrial septal defects

被引:56
作者
Rickers, C
Jerosch-Herold, M
Hu, XD
Murthy, N
Wang, XE
Kong, HF
Seethamraju, RT
Weil, J
Wilke, NM
机构
[1] Univ Minnesota, Dept Radiol, Minneapolis, MN 55455 USA
[2] Univ Hamburg, Hosp Eppendorf, D-20246 Hamburg, Germany
关键词
pediatrics; heart defects; congenital; magnetic resonance imaging;
D O I
10.1161/01.CIR.0000039343.95540.CF
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Recent developments in cardiac MRI have extended the potential spectrum of diagnostic and interventional applications. The purpose of this study was to test the ability of MRI to perform transcatheter closures of secundum type atrial septal defects (ASD) and to assess ASD size and changes in right cardiac chamber volumes in the same investigation. Methods and Results-In 7 domestic swine (body weight, 38+/-13 kg), an ASD (Q(p):Q(s)=1.7+/-0.2) was created percutaneously by balloon dilation of the fossa ovalis. The ASD was imaged and sized by both conventional radiography and MRI. High-resolution MRI of the ASD diameters correlated well with postmortem examination (r=0.97). Under real-time MR fluoroscopy, the introducer sheath was tracked toward the left atrium with the use of novel miniature MR guide wires. The defect was then closed with an Amplatzer Septal Occluder. In all animals, it was possible to track and interactively control the position of the guide wire within the vessels and the heart, including the successful deployment of the Amplatzer Septal Occluder. Right atrial and ventricular volumes were calculated before and after the intervention by using cine-MRI. Both volumes were found to be significantly reduced after ASD closure (P<0.005). Conclusions-These in vivo studies demonstrate that catheter tracking and ASD device closure can be performed under real-time MRI guidance with the use of intravascular antenna guide wires. High-resolution imaging allows accurate determination of ASD size before the intervention, and immediate treatment effects such as changes in right cardiac volumes can also be measured.
引用
收藏
页码:132 / 138
页数:7
相关论文
共 29 条
[1]   High resolution intravascular MRI and MRS by using a catheter receiver coil [J].
Atalar, E ;
Bottomley, PA ;
Ocali, O ;
Correia, LCL ;
Kelemen, MD ;
Lima, JAC ;
Zerhouni, EA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (04) :596-605
[2]   Editorial: Transcatheter closure of atrial septal defects [J].
Austin, EH .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 120 (06) :1032-1033
[3]   Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions [J].
Bakker, CJ ;
Hoogeveen, RM ;
Weber, J ;
vanVaals, JJ ;
Viergever, MA ;
Mali, WP .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) :816-820
[4]   NUCLEAR MAGNETIC-RESONANCE (NMR) INVIVO STUDIES - KNOWN THRESHOLDS FOR HEALTH-EFFECTS [J].
BUDINGER, TF .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1981, 5 (06) :800-811
[5]  
Gu XP, 1999, CATHETER CARDIO INTE, V46, P51, DOI 10.1002/(SICI)1522-726X(199901)46:1<51::AID-CCD14>3.0.CO
[6]  
2-M
[7]   IMAGING AND SIZING OF ATRIAL SEPTAL-DEFECTS BY MAGNETIC-RESONANCE [J].
HOLMVANG, G ;
PALACIOS, IF ;
VLAHAKES, GJ ;
DINSMORE, RE ;
MILLER, SW ;
LIBERTHSON, RR ;
BLOCK, PC ;
BALLEN, B ;
BRADY, TJ ;
KANTOR, HL .
CIRCULATION, 1995, 92 (12) :3473-3480
[8]  
HU X, 1996, MAGNET RESON MED, V31, P691
[9]   Real-time interactive MRI on a conventional scanner [J].
Kerr, AB ;
Pauly, JM ;
Hu, BS ;
Li, KC ;
Hardy, CJ ;
Meyer, CH ;
Macovski, A ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (03) :355-367
[10]   Target definition and trajectory optimization for interactive MR-guided biopsies of brain tumors in an open configuration MRI system [J].
Kollias, SS ;
Bernays, R ;
Marugg, RA ;
Romanowski, B ;
Yonekawa, Y ;
Valavanis, A .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :143-159