A new readout system for bolometers with improved low frequency stability

被引:28
作者
Gaertner, S
Benoit, A
Lamarre, JM
Giard, M
Bret, JL
Chabaud, JP
Desert, FX
Faure, JP
Jegoudez, G
Lande, J
Leblanc, J
Lepeltier, JP
Narbonne, J
Piat, M
Pons, R
Serra, G
Simiand, G
机构
[1] CTR RECH TRES BASSES TEMP,F-38042 GRENOBLE 09,FRANCE
[2] UNIV PARIS 11,INST ASTROPHYS SPATIALE,F-91405 ORSAY,FRANCE
来源
ASTRONOMY & ASTROPHYSICS SUPPLEMENT SERIES | 1997年 / 126卷 / 01期
关键词
instrumentation; detectors; photometer; methods; data analysis; numerical;
D O I
10.1051/aas:1997256
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
A new readout electronic system for bolometers is presented in this paper. The bolometer resistance is measured in a bridge with a capacitive load, using a periodic square wave bias current. The bias voltages at both ends of the bridge are balanced in order to keep the middle point around zero. Only changes around this zero value are amplified and detected synchronously with the bias signal. These features shift the measurement frequency out of the electrical low frequency noises (JFETs), and reduces the dynamics required from the amplification chain. The bias voltages are fully controlled by computer, and the lock-in detection is digital. This readout electronic has many advantages over previous ones. In particular, it proved to be able to read the total power of the radiation reaching the bolometer, and to perform measurements down to low frequencies (less than or equal to 0.1 Hz) without significant additional noise. These features open new observation strategies such as full sky scanning to bolometer instruments on board future submillimetre space projects having high thermal background levels (warm telescope). The different steps of the development of this new readout electronics on the ground-based Diabolo experiment are described, the performances reached are discussed, and a version suitable for the readout of the arrays of bolometers on the COBRAS/SAMBA satellite mission is presented.
引用
收藏
页码:151 / 160
页数:10
相关论文
共 15 条
[11]   BOLOMETERS - ULTIMATE SENSITIVITY, OPTIMIZATION, AND AMPLIFIER COUPLING [J].
MATHER, JC .
APPLIED OPTICS, 1984, 23 (04) :584-588
[12]   BOLOMETER NOISE - NON-EQUILIBRIUM THEORY [J].
MATHER, JC .
APPLIED OPTICS, 1982, 21 (06) :1125-1129
[13]  
MURAKAMI H, 1996, IN PRESS PASJ
[14]   AN AC BRIDGE READOUT FOR BOLOMETRIC DETECTORS [J].
RIEKE, FM ;
LANGE, AE ;
BEEMAN, JW ;
HALLER, EE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :946-949
[15]   IMPROVED LOW-FREQUENCY STABILITY OF BOLOMETRIC DETECTORS [J].
WILBANKS, T ;
DEVLIN, M ;
LANGE, AE ;
SATO, S ;
BEEMAN, JW ;
HALLER, EE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :566-572