Biomechanical evaluation of vertebroplasty and kyphoplasty with polymethyl methacrylate or calcium phosphate cement under cyclic loading

被引:126
作者
Wilke, Hans-Joachim
Mehnert, Ulrich
Claes, Lutz E.
Bierschneider, Michael M.
Jaksche, Hans
Boszczyk, Bronek M.
机构
[1] Univ Ulm, Inst Orthopaed Res & Biomech, D-89081 Ulm, Germany
[2] Berufsgenossenschaftliche Unfallklin, Murnau, Germany
[3] Univ Bern, Dept Orthopaed Surg, Spine Serv, Bern, Switzerland
关键词
vertebroplasty; kyphoplasty; biomechanics; cyclic loading; polymethylmethacrylate; calcium phosphate;
D O I
10.1097/01.brs.0000248423.28511.44
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. We developed a new method to simulating in vivo dynamic loading as closely as possible, which allows comparison of kyphoplasty and vertebroplasty, as well as augmentation materials. Objective. Special interest was given to calcium phosphate cement, which might fail due to its brittleness. Summary of Background Data. Vertebroplasty and kyphoplasty are, with limitations, 2 promising alternative techniques to augment osteoporotic vertebrae with polymethyl methacrylate or calcium phosphate cements. However, little is known about the fatigue characteristics of the treated vertebrae under cyclic loading. Methods. Twenty-four intact, osteoporotic bi-segmental human specimens were divided into 4 groups: (1) vertebroplasty with polymethyl methacrylate, (2) kyphoplasty with polymethyl methacrylate, (3) kyphoplasty with calcium phosphate cement, and (4) untreated control group. After augmentation of the middle vertebrae, all specimens underwent 100,000 cycles of eccentric loading during which the specimen revolved around its longitudinal axis. Pre-loading and post-loading radiographs, and subsidence measurements at different sites of the vertebrae were taken. The overall height was additionally determined every 20,000 cycles in the material testing machine. Finally, the specimens were cryosectioned to examine the cements. Results. Loss of height progressed with strong individual differences in all groups, with an increasing number of load cycles up to median values of 2.8 mm for both augmented groups and 4.2 mm for the nonaugmented group. At the center of the upper endplate, subsidence in kyphoplasty was greater than in vertebroplasty, with little differences with respect to the kind of cement. The cryosections did not show any signs of fatigue in the polymethyl methacrylate, but small cracks were in the calcium phosphate. Conclusions. Vertebroplasty and kyphoplasty seem to be equivalent methods in strengthening osteoporotic vertebrae. However, these results cannot be transferred to the treatment of fractures with these methods. A "physiologic" loading situation was achieved by complex motion, including all combinations of flexion/extension with lateral bending during eccentric cyclic loading.
引用
收藏
页码:2934 / 2941
页数:8
相关论文
共 25 条
[1]   The use of an injectable, biodegradable calcium phosphate bone substitute for the prophylactic augmentation of osteoporotic vertebrae and the management of vertebral compression fractures [J].
Bai, B ;
Jazrawi, LM ;
Kummer, FJ ;
Spivak, JM .
SPINE, 1999, 24 (15) :1521-1526
[2]   Percutaneous vertebroplasty for pain relief and spinal stabilization [J].
Barr, JD ;
Barr, MS ;
Lemley, TJ ;
McCann, RM .
SPINE, 2000, 25 (08) :923-928
[3]   The biomechanics of vertebroplasty - The effect of cement volume on mechanical behavior [J].
Belkoff, SM ;
Mathis, JM ;
Jasper, LE ;
Deramond, H .
SPINE, 2001, 26 (14) :1537-1541
[4]   An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of compression fracture [J].
Belkoff, SM ;
Mathis, JM ;
Fenton, DC ;
Scribner, RM ;
Reiley, ME ;
Talmadge, K .
SPINE, 2001, 26 (02) :151-156
[5]  
Belkoff SM, 2002, AM J NEURORADIOL, V23, P1647
[6]   Adjacent vertebral failure after vertebroplasty - A biomechanical investigation [J].
Berlemann, U ;
Ferguson, SJ ;
Nolte, LP ;
Hein, PF .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (05) :748-752
[7]   Physical and chemical aspects of calcium phosphates used in spinal surgery [J].
Bohner, M .
EUROPEAN SPINE JOURNAL, 2001, 10 (Suppl 2) :S114-S121
[8]  
Cortet B, 1999, J RHEUMATOL, V26, P2222
[9]  
DERAMOND H, 1999, BONE S, V25, P17
[10]   New technologies in spine - Kyphoplasty and vertebrosplasty for the treatment of painful osteoporotic compression fractures [J].
Garfin, SR ;
Yuan, HA ;
Reiley, MA .
SPINE, 2001, 26 (14) :1511-1515