Study of the potentiometric response towards sodium ions of Na0.44-xMnO2 for the development of selective sodium ion sensors

被引:79
作者
Sauvage, F. [1 ]
Baudrin, E. [1 ]
Tarascon, J. -M. [1 ]
机构
[1] Univ Picardie, CNRS, UMR 6007, Lab React & Chim Solides, F-80039 Amiens, France
关键词
potentiometric ion sensors; insertion materials; manganese oxide; Na0.44MnO2;
D O I
10.1016/j.snb.2006.03.024
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Single phase Na0.44MnO2 powder was synthetized by a classical solid-state reaction. Using a plastic technology, we studied the potentiometric response versus Na+ ions in aqueous medium. A Nernstian-like response (55 mV/dec) was obtained for concentrations ranging from 1 M to 2 x 10(-4) M with a quite good selectivity towards Li+, K+, Mg2+ and Ca2+. Through an ageing study of this material in aqueous solution, we clearly identified the sensing material as being the Na0.33MnO2 phase. Furthermore, we proved that the sensing activity of this compound is routed in its ability to reversibly insert/de-insert Na+ ions. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:638 / 644
页数:7
相关论文
共 29 条
[1]   Response behavior of sodium-selective electrodes modified by surface attachment of the anticoagulant polysaccharides heparin and chondroitin sulfate [J].
Badr, IHA ;
Feiler, J ;
Bachas, LG .
TALANTA, 2005, 65 (01) :261-266
[2]  
BAKKER E, 1997, TRENDS ANAL CHEM, V16
[3]   A sodium ion sensor based on a covalently-linked aminorhodamine B-calix[4]arene chromoionophore [J].
Benco, JS ;
Nienaber, HA ;
McGimpsey, WG .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 85 (1-2) :126-130
[4]  
CANEIRO A, 1991, ANAL CHEM, P63
[5]   Microsensor based on NASICON for the analysis of Na+ in the gingival fluid [J].
Caravel, G ;
Poignet, N ;
Siebert, E ;
Fabry, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 76 (1-3) :506-511
[6]   A sodium ion selective optode based on fluorescein octadecyl ether octadecyl ester and application in beverage and urine assay [J].
Chan, WH ;
Lee, AWM ;
Lam, YS ;
Lu, JZ .
MICROCHEMICAL JOURNAL, 2002, 72 (02) :201-207
[7]   THE SODIUM RESPONSIVE TUNGSTEN BRONZE ELECTRODE - AN ELECTRODE FOR SODIUM-ION ANALYSIS OF DNA SOLUTIONS [J].
DOBSON, JV ;
COMER, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1987, 220 (02) :225-234
[8]   Lithium insertion processes of orthorhombic NaxMnO2-based electrode materials [J].
Doeff, MM ;
Richardson, TJ ;
Kepley, L .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1996, 143 (08) :2507-2516
[9]   Synthesis and characterization of a copper-substituted manganese oxide with the Na0.44MnO2 structure [J].
Doeff, MM ;
Richardson, TJ ;
Hollingsworth, J ;
Yuan, CW ;
Gonzales, M .
JOURNAL OF POWER SOURCES, 2002, 112 (01) :294-297
[10]  
DOEFF MM, 1994, J ELECTROCHEM SLC, V141