Optimization of chemiluminescent nitrogen detection for packed column supercritical fluid chromatography with methanol-modified CO2

被引:11
作者
Combs, MT
AshrafKhorassani, M
Taylor, LT
Fujinari, EM
机构
[1] VIRGINIA POLYTECH INST & STATE UNIV, DEPT CHEM, BLACKSBURG, VA 24061 USA
[2] ANTEK INSTRUMENTS INC, HOUSTON, TX 77090 USA
关键词
D O I
10.1021/ac961138w
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A chemiluminescent nitrogen detector was optimized for packed-column supercritical fluid chromatography. Methanol modifier concentrations of 15% or less, an ozone now of 5.8 mL/min, and a decompressed CO2 now between 240 and 310 mL/min were found to exhibit a maximum sensitivity of 5 ng for sulfamethazine (1 ng of N). The addition of a membrane drier to the ''optimized'' system further decreased the minimum detectable quantity (MDQ) to 0.5 ng (0.1 ng of N). In addition, by using a microbore column (2 mm i.d.) instead of an analytical scale column, the postcolumn decompressed now split could be eliminated, further reducing the MDQ to 0.125 ng of sulfamethazine (0.025 ng of N).
引用
收藏
页码:3044 / 3048
页数:5
相关论文
共 12 条
[1]  
BENN SM, 1993, DEV FOOD SCI, V32, P65
[2]  
Bizanek R, 1996, PEPTIDE RES, V9, P40
[3]  
BRITTEN AJ, 1989, RES DEV, V31, P76
[4]  
CHAWLA B, IN PRESS J CHROMATOG
[5]  
COURTHAUDON LO, 1991, LC GC-MAG SEP SCI, V9, P732
[6]   NITROGEN-SPECIFIC DETECTION OF PEPTIDES IN LIQUID-CHROMATOGRAPHY WITH A CHEMILUMINESCENT NITROGEN DETECTOR [J].
FUJINARI, EM ;
MANES, JD .
JOURNAL OF CHROMATOGRAPHY A, 1994, 676 (01) :113-120
[7]   NITROGEN-SPECIFIC LIQUID-CHROMATOGRAPHY DETECTOR BASED ON CHEMILUMINESCENCE - APPLICATION TO THE ANALYSIS OF AMMONIUM NITROGEN IN WASTE-WATER [J].
FUJINARI, EM ;
COURTHAUDON, LO .
JOURNAL OF CHROMATOGRAPHY, 1992, 592 (1-2) :209-214
[8]   Peptide content-determination of crude synthetic peptides by reversed-phase liquid chromatography and nitrogen-specific detection with a chemiluminescent nitrogen detector [J].
Fujinari, EM ;
Manes, JD ;
Bizanek, R .
JOURNAL OF CHROMATOGRAPHY A, 1996, 743 (01) :85-89
[9]   Feasibility of supercritical fluid chromatography chemiluminescent nitrogen detection with open tubular columns [J].
Shi, H ;
Strode, JTB ;
Taylor, LT ;
Fujinari, EM .
JOURNAL OF CHROMATOGRAPHY A, 1996, 734 (02) :303-310
[10]  
SHI H, IN PRESS J CHROMATOG