Dosimetric characteristics of the InterSource103 palladium brachytherapy source

被引:44
作者
Meigooni, AS [1 ]
Sowards, K [1 ]
Soldano, M [1 ]
机构
[1] Univ Kentucky, Albert B Chandler Med Ctr, Dept Radiat Med, Lexington, KY 40536 USA
关键词
Pd-103; brachytherapy; dosimetry; thermoluminescent dosimeter;
D O I
10.1118/1.598991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Pd-103 brachytherapy sources are being used for interstitial implants in tumor sites such as the prostate. Recently, the InterSource(103) palladium source has been introduced, which has a design different from that of other sources presently on the market. Dosimetric characteristics (i.e., dose rate constant, radial dose function, and anisotropy function) of this source were experimentally and theoretically determined following the AAPM Task Group 43 (TG-43) recommendations and were related to the 1999 NIST calibration assigned to this source [S-k, 99std]. Measurements were performed in a solid water phantom using LiF thermoluminescent dosimeters. The measured data was compared with Monte Carlo simulations performed in solid water using the PTRAN code. The calculations were then performed in liquid water to obtain the dosimetric information for clinical applications as per TG-43 recommendation. The results indicated that the dose rate constant, Lambda, of the InterSource(103) palladium source was 0.664 +/- 5% cGy/h/U using TLD measurements and 0.660 +/- 3% cGy/h/U using Monte Carte simulations in solid water. The calculated value in liquid water was found to he 0.696 +/- 3% cGy/h/U. The radial dose function, g(r), of the new Pd-103 source was measured at distances ranging from 0.5 to 10 cm using LiF TLD in solid water phantom material. The Monte Carlo simulations were performed at distances ranging from 0.1 to 10 cm from the source center in solid water and liquid water. The anisotropy function, F(r, theta), was measured at distances of 2, 3, 5, and 7 cm from the source center and calculated at distances of 0.5, 1, 2, 3, 5, and 7 cm from the source center. Complete dosimetric data are described in this paper. Per TG-43, the values reported in water should be used for clinical treatment planning systems. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)03205-3].
引用
收藏
页码:1093 / 1100
页数:8
相关论文
共 23 条
[1]
PROSTATE-SPECIFIC ANTIGEN BASED DISEASE-CONTROL FOLLOWING ULTRASOUND-GUIDED (125)IODINE IMPLANTATION FOR STAGE T1/T2 PROSTATIC-CARCINOMA [J].
BLASKO, JC ;
WALLNER, K ;
GRIMM, PD ;
RAGDE, H .
JOURNAL OF UROLOGY, 1995, 154 (03) :1096-1099
[2]
THERMOLUMINESCENT DOSIMETRY FOR PD-103 SEEDS (MODEL 200) IN SOLID WATER PHANTOM [J].
CHIUTSAO, ST ;
ANDERSON, LL .
MEDICAL PHYSICS, 1991, 18 (03) :449-452
[3]
Hubbell J. H., 1995, 5632 NISTIR
[4]
VOLUME-BASED GEOMETRIC MODELING FOR RADIATION TRANSPORT CALCULATIONS [J].
LI, ZF ;
WILLIAMSON, JF .
MEDICAL PHYSICS, 1992, 19 (03) :667-677
[5]
INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS [J].
MEIGOONI, AS ;
MISHRA, V ;
PANTH, H ;
WILLIAMSON, J .
MEDICAL PHYSICS, 1995, 22 (05) :555-561
[6]
Meigooni AS, 1995, BRACHYTHERAPY PHYSIC, P210
[7]
MEIGOONI AS, 1990, ENDOCURIETHER HYPERT, V6, P107
[8]
PHOTON ENERGY-DEPENDENCE OF THE SENSITIVITY OF RADIOCHROMIC FILM AND COMPARISON WITH SILVER-HALIDE FILM AND LIF TLDS USED FOR BRACHYTHERAPY DOSIMETRY [J].
MUENCH, PJ ;
MEIGOONI, AS ;
NATH, R ;
MCLAUGHLIN, WL .
MEDICAL PHYSICS, 1991, 18 (04) :769-775
[9]
DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[10]
NCRP, 1985, 58 NCRP